Instruction d'utilisation Boca Research Muon Liquid Handling System None

Instruction d'utilisation pour le dispositif Boca Research Muon Liquid Handling System None

Dispositif: Boca Research Muon Liquid Handling System None
Catégorie: Pompe de chauffage
Fabricant: Boca Research
Dimension: 0.45 MB
Date d'addition: 5/22/2014
Nombre des pages: 14
Imprimez l'instruction

Téléchargez

Comment utiliser?

Notre objectif consiste à vous assurer le plus rapide accès au contenu de l'instruction d'utilisation du dispositif Boca Research Muon Liquid Handling System None. En utilisant l'aperçu en ligne vous pouvez parcourir le sommaire et passer à la page où vous trouverez la résolution de votre problème avec Boca Research Muon Liquid Handling System None.

Pour votre confort

Si regarder l'instruction Boca Research Muon Liquid Handling System None directement sur la page n'est pas confortable pour vous, vous pouvez profiter de deux solutions possibles:

  • Exploration au mode de plein écran - pour analyser l'instruction d'une manière commode (sans l'avoir téléchargée sur votre ordinateur) vous pouvez utiliser le mode d'exploration en plein écran. Pour démarrer l'exploration de l'instruction Boca Research Muon Liquid Handling System None en plein écran, utilisez le bouton Plein écran.
  • Téléchargement sur l'ordinateur - vous pouvez aussi télécharger l'instruction Boca Research Muon Liquid Handling System None sur votre ordinateur et la sauvegarder dans vos ressources. Si vous n'avez pas envie de gaspiller votre espace disque, vous pouvez toujours télécharger cette instruction du ManualsBase dans l'avenir.
Boca Research Muon Liquid Handling System None Manuel d'utilisation - Online PDF
Advertisement
« Page 1 of 14 »
Advertisement
Version imprimée

Beaucoup de gens préfèrent de lire les documents dans la version imprimée et non pas sur l'écran. L'option d'impression de l'instruction est aussi prévue et vous pouvez en profiter en cliquant le lien ci-dessus - Imprimez l'instruction. Il n'est pas nécessaire d'imprimer toute l'instruction Boca Research Muon Liquid Handling System None - on peut choisir des pages. Economisez le papier.

Résumés

Vous trouverez ci-dessous les annonces des contenus qui se trouvent sur les pages suivantes de l'instruction de Boca Research Muon Liquid Handling System None. Si vous voulez parcourir rapidement le contenu des pages suivantes de l'instruction, vous pouvez en profiter.

Résumés du contenu
Résumé du contenu de la page N° 1



Muon Liquid Handling System
User Guide




C. Johnson, S.P. Cottrell et al
Version 0.2












- 1 -

Résumé du contenu de la page N° 2

Contents 1 Getting Started 3 1.1 Layout of the liquid handling system 3 1.2 Layout of the in situ sample stick 4 1.3 Layout of the pump 4 2 General procedures 6 2.1 Sample loading 6 2.1.1 Loading Vessel 1 6 2.1.2 Loading Vessel 2 6 2.2 Evacuation of the system 6 2.2.1 Removing air from Vessel 1 7 2.2.2 Removing air from Vessel 2 7 2.2.3 Removing air from Vessel 3 7 2.2.4 Evacuation of the sample cell 7

Résumé du contenu de la page N° 3

1 Getting Started This manual describes the liquid handling system and the in situ sample cell as used on the DEVA instrument with the “RF” spectrometer, for either normal muon spins relaxation or RF resonance experiments. 1.1 Layout of the liquid handling system The Muon Liquid Handling System has been designed to facilitate the in situ degassing and transfer of samples into and out of the sample cell. The main features of the liquid handling system are shown in Figure 1. Inl

Résumé du contenu de la page N° 4

1.2 Layout of the in situ sample stick The liquid sample stick is designed to fit into the DEVA flow cryostat, details of which can be found in the DEVA manual. It consists of a shapol target cell 30 mm x 30 mm with a mylar window upon which may be mounted an RF coil. Two stainless steel capillaries provide a means of flowing liquid into and out of the cell along with feed throughs for an RF excitation signal and a pick-up coil. Figure 2 The liquid-sample cell mounted on a sampl

Résumé du contenu de la page N° 5

Roughing/ Turbo Off/Run switch switch Pressure reading Turbo control panel Turbo Pump Rotary Pump Figure 2 The layout of the front panel on the vacuum pump used with the muon liquid handling system. - 5 -

Résumé du contenu de la page N° 6

2 General procedures 2.1 Sample loading At this point, the system will be open to the atmosphere, ensure that valve V2 is closed to prevent air condensation in the cold trap. 2.1.1 Loading Vessel 1 • Open the tap on the right-hand side of Vessel 1. • Open tap A and valve V11. • The liquid should be poured into the system very slowly to avoid liquid leaving the vessel via the side tap. • When loading is complete, close the side tap, tap A and valve V11. 2.1.2 Loading Vessel 2

Résumé du contenu de la page N° 7

2.2.1 Removing air from Vessel 1 Once the system has been pressurised, tap A on Vessel 1 can be opened. It is possible to bubbled helium gas through the sample liquid against the non-return valve, V13. If this is required, first ensure that valves V5, V7 and V8 are closed. Open valves V10 and V6, finally opening tap B slowly in order to regulate the flow. To fully evacuate the system and Vessel 1, the sample liquid must be frozen by raising a dewar of liquid nitrogen around it. Wh

Résumé du contenu de la page N° 8

2.3 Freeze-Pump-Thaw cycle Using this system it is possible to operate freeze-pump-thaw cycles on sample liquids held in any of the three vessels. Procedures appropriate to each vessel are outlined below. 2.3.1 Sample in Vessel 1 • All taps and valves should be closed. • Open valve V2 and tap A on Vessel 1, and freeze the sample liquid by slowly raising a dewar of liquid nitrogen around it. • When it has been determined that the liquid has completely frozen switch the pump to ro

Résumé du contenu de la page N° 9

2.3.3 Sample in Vessel 3 • All taps and valves should be closed. • Open valves V2 and V3 along with tap C on Vessel 3 and freeze the sample liquid by slowly raising a dewar of liquid nitrogen around it. • When it has been determined that the liquid has completely frozen switch the pump to roughing mode and open valve V1. • Switch the pump to turbo mode and leave for several minutes. The pressure within the system can be read from the baratron gauge on the pump. • When sufficient pump

Résumé du contenu de la page N° 10

2.4.2 Vessel 3 to sample cell The transfer of liquid from Vessel 3 to the sample cell is brought about by the difference in pressure between that in the Vessel and a vacuum in the cell. However, in order to control the transfer of liquid the pressure difference should be as small as possible but not so low that the liquid starts to boil. • Switch the pump to its roughing mode. • Open values V2 and V3 along with tap C and carefully open valve V1 until a suitable pressure is obtained (

Résumé du contenu de la page N° 11

2.4.4 Recovery of liquid in sample cell to Vessel 1 The liquid in the sample cell cannot be recovered to Vessel 1 using the liquid handling rig as presently configured. 2.4.5 Recovery of liquid in sample cell to Vessel 2 The liquid in the sample cell can be recovered to Vessel 2 using the following procedure. • Vessel 2 should be under vacuum along with the manifold between valves V5 and V6 prior to the recovery of the liquid from the sample cell. If it is not, switch the pump t

Résumé du contenu de la page N° 12

• All the liquid may not return in a single pass and repeating the evacuation procedure may be required. However, some liquid may be present in the system and care should be taken during evacuation that the trap does not become blocked. The liquid may also be encouraged to leave the sample cell by using a small amount of gas. • Close valve V3 and open valve V10 to allow helium gas into the lower part of the system. Close valve V10 and carefully open valve V5 to allow the gas to pass th

Résumé du contenu de la page N° 13

3 Example Experiment The paramagnetic signal in liquid n-hexane is difficult to observe because the muonium polarisation is found to decay on a microsecond time scale, which is further reduced by the presence of dissolved oxygen. A series of commissioning experiments were carried out using n-hexane in order to gauge the ability of the liquid handling rig to degas the liquid. This experiment was carried out using a detachable liquid-sample cell. At the start of the experiment, 100 m

Résumé du contenu de la page N° 14

13.5 13.0 a) b) 13.0 12.5 12.5 12.0 12.0 11.5 11.5 11.0 11.0 0.0 0.5 1.0 1.5 2.0 2.5 3.0 0.0 0.5 1.0 1.5 2.0 2.5 3.0 Time (µs) Time (µs) c) 13.5 13.0 12.5 12.0 11.5 11.0 0.0 0.5 1.0 1.5 2.0 2.5 3.0 Time (µs) Figure 1 Muonium precession signal at 2G in n-hexane after a) 4, b) 7 and c) 10 freeze- pump-thaw cycles. For the second experiment, all of the n-hexane used in the first experiment was recovered into a sample bottle and dried using a small piece of clean sodium. The liquid was c


Instructions pareilles
# Instruction d'utilisation Catégorie Téléchargez
1 Edelbrock 15005 Manuel d'utilisation Pompe de chauffage 0
2 A.O. Smith AH Manuel d'utilisation Pompe de chauffage 2
3 ACS 882.00217.00 Manuel d'utilisation Pompe de chauffage 1
4 A.O. Smith AHD 290 Manuel d'utilisation Pompe de chauffage 3
5 AC International SELF-PRIMING CENTRIFUGAL PUMPS Manuel d'utilisation Pompe de chauffage 5
6 Acnodes Q5RF SERIES Manuel d'utilisation Pompe de chauffage 2
7 Amana 10730418 Manuel d'utilisation Pompe de chauffage 1
8 Allied Air Enterprises (2 Manuel d'utilisation Pompe de chauffage 0
9 Allied Air Enterprises 15) Manuel d'utilisation Pompe de chauffage 0
10 Allied Air Enterprises 4HP16LT Manuel d'utilisation Pompe de chauffage 2
11 AC International 8200 Series Manuel d'utilisation Pompe de chauffage 1
12 American Standard TWE060B***E Manuel d'utilisation Pompe de chauffage 2
13 Allied Air Enterprises 4)PHP(13 Manuel d'utilisation Pompe de chauffage 1
14 American Standard TWE120A***F Manuel d'utilisation Pompe de chauffage 0
15 American Standard TWE090B***E Manuel d'utilisation Pompe de chauffage 0