Teledyne OXYGEN DETECTOR 319R user manual

User manual for the device Teledyne OXYGEN DETECTOR 319R

Device: Teledyne OXYGEN DETECTOR 319R
Category: Oxygen Equipment
Manufacturer: Teledyne
Size: 0.07 MB
Added : 8/27/2013
Number of pages: 23
Print the manual

Download

How to use this site?

Our goal is to provide you with a quick access to the content of the user manual for Teledyne OXYGEN DETECTOR 319R. Using the online preview, you can quickly view the contents and go to the page where you will find the solution to your problem with Teledyne OXYGEN DETECTOR 319R.

For your convenience

If looking through the Teledyne OXYGEN DETECTOR 319R user manual directly on this website is not convenient for you, there are two possible solutions:

  • Full Screen Viewing - to easily view the user manual (without downloading it to your computer), you can use full-screen viewing mode. To start viewing the user manual Teledyne OXYGEN DETECTOR 319R on full screen, use the button Fullscreen.
  • Downloading to your computer - You can also download the user manual Teledyne OXYGEN DETECTOR 319R to your computer and keep it in your files. However, if you do not want to take up too much of your disk space, you can always download it in the future from ManualsBase.
Teledyne OXYGEN DETECTOR 319R User manual - Online PDF
Advertisement
« Page 1 of 23 »
Advertisement
Print version

Many people prefer to read the documents not on the screen, but in the printed version. The option to print the manual has also been provided, and you can use it by clicking the link above - Print the manual. You do not have to print the entire manual Teledyne OXYGEN DETECTOR 319R but the selected pages only. paper.

Summaries

Below you will find previews of the content of the user manuals presented on the following pages to Teledyne OXYGEN DETECTOR 319R. If you want to quickly view the content of pages found on the following pages of the manual, you can use them.

Abstracts of contents
Summary of the content on the page No. 1

MODEL
319R
OXYGEN DETECTOR
INSTRUCTION MANUAL
P/N M59608
ECO# 95-416
7/10/95
TELEDYNE BROWN EGINEERING
Analytical Instruments

Summary of the content on the page No. 2

Copyright © 1993 Teledyne Brown Engineering Analytical Instruments All Rights Reserved. No part of this manual may be reproduced, transmitted, transcribed, stored in a retrieval system, or translated into any other language or computer language in whole or in part, in any form or by any means, whether it be electronic, mechanical, magnetic, optical, manual, or otherwise, without the prior written consent of Teledyne Brown Egineering Analytical Instruments (TBE), 16830 Chestnut Street, City of In

Summary of the content on the page No. 3

Table of Contents 1.0 Introduction Main Features ............................................................................................... 1 Micro-Fuel Cell ............................................................................................ 2 Z-2 Cell Warranty Conditions ...................................................................... 2 2.0 Installation & Operation Precautions....................................................................................................

Summary of the content on the page No. 4

TELEDYNE BROWN EGINEERING Analytical Instruments

Summary of the content on the page No. 5

319R Introduction 1.0 Introduction The Teledyne Brown Engineering Analytical Instruments (TBE) Model 319R Oxygen Detector measures the percent of oxygen in a wide range of gases and gas mixtures. Modular and compact design provide outstanding performance and reliable on-line monitoring and make the 319R virtually maintenance-free. A special range on the 319R enables you to calibrate in ambient air. The Micro-Fuel Cell produces an output that corresponds linearly to the oxygen concentration; the

Summary of the content on the page No. 6

1.0 Introduction 319R 4. Proven Sensor. The 319R uses the Class Z-2 Micro-Fuel Cell, warranted for 6 months, with a life expectancy of 6 months. 5. Analog Voltage Output. The Model 319R provides customer connections to external devices, such as computers, recorders, alarms, etc., at 0–10 VDC, negative ground, plus contact closures for range identifica- tion. Custom voltage outputs, either grounded or isolated, are available as options. 6. Analog Current Output. The Model 319R has a 4–20 mA curre

Summary of the content on the page No. 7

319R Installation & Operation 2.0 Installation & Operation Precautions 1. Read this chapter in its entirety before installing the unit. 2. Upon receipt of the instrument, inspect the entire unit for damage. Notify the shipper if any damage is found. Check the unit, probe, and included accessories for broken or loose parts. 3. The sample must be free of entrained solids and water. A high humidity sample is ideal since it will prevent water loss from the cell electrolyte. 4. The Micro-Fuel Cell s

Summary of the content on the page No. 8

2.0 Installation & Operation 319R The analysis unit should be installed in the process at a maximum distance of 6 feet away from the main unit, depending upon the level of electromagnetic noise in the operating environment. Analysis Unit Connection The receptacle for the analysis unit cable is located in the lower left- hand corner of the rear panel. The 6-pin Minifit fitting is keyed to fit only one way in the receptacle; do not force it in. The other end of the cable is made of four separate

Summary of the content on the page No. 9

319R Installation & Operation 2.0 Alarm Terminal Connection SPDT relay contacts are provided for each of the 3 alarms. There are 9 terminals available on the back panel of the system for alarm circuit connection. The explanation for the relay terminals abbreviations is as follows: NC= Normally closed C = Common NO= Normally open The alarm relay circuits are designed for failsafe operation, meaning that the relays are activated (energized) during normal operation. If power fails, or if an alarm

Summary of the content on the page No. 10

2.0 Installation & Operation 319R Setting Alarms The alarms can be adjusted over the full scale range, and they are set as a percent of full scale. Thus, an alarm set to indicate 192 on the display while on the 200 PPM O range will activate at 192 PPM O on the 200 2 2 PPM O range and at 1920 O on the 2000 O range (96% of full scale). To 2 2 2 set the alarms: 1. Place the range switch located below the front panel display in the 2000 PPM range position (or highest range available). 2. Place the

Summary of the content on the page No. 11

319R Installation & Operation 2.0 Calibration To calibrate the 319R: 1. Place the power switch in the ON position. 2. Allow the LCD reading to stabilize. 3. Expose the sensor to ambient air (20.9%) or instrument grade air. If the analyzer goes overrange, the display will go blank and the LED to the left of the display will blink. Hold down the SPAN button until the LED stops blinking. 4. While holding down the SPAN AT 20.9% button, adjust the potentiometer to the right of the button until the d

Summary of the content on the page No. 12

2.0 Installation & Operation 319R TELEDYNE BROWN ENGINEERING 8 Analytical Instruments

Summary of the content on the page No. 13

319R Maintenance 3.0 Maintenance Replacing the Fuse If the unit will not turn on when plugged in and switched ON (via the power switch), or you suspect a blown fuse, replace the fuse using the following procedure: 1. Disconnect the AC power and place the power switch located on the rear panel in the OFF position. 2. The fuse is located next to the power cord receptacle in the upper left-hand corner of the rear panel. Remove the power cord from the receptacle. 3. Insert a small flat-blade screwd

Summary of the content on the page No. 14

3.0 Maintenance 319R 5. Replace the fuse holder into its receptacle, pushing in firmly until it clicks. In units with DC power, the fuse is located on the rear panel above the ON/OFF switch. Open the fuse holder by unscrewing and removing the knob marked FUSE. The fuse will be located inside the knob. Replacing the Sensor The characteristics of the Micro-Fuel Cell are similar to those of a mercury battery in that both provide an almost constant output through their useful life, and then fall of

Summary of the content on the page No. 15

319R Maintenance 3.0 Warranty Conditions The Class Z-2 cell used in the 319R is warranted for six (6) months of service from the date of shipment. Customers having warranty claims must return the cell in question to the factory for evaluation. If it is determined that failure is due to faulty material or workmanship, the cell will be replaced free of charge. CAUTION: Evidence of tampering or abuse will render the warranty null and void. TELEDYNE BROWN ENGINEERING 11 Analytical Instruments

Summary of the content on the page No. 16

3.0 Maintenance 319R TELEDYNE BROWN ENGINEERING 12 Analytical Instruments

Summary of the content on the page No. 17

319R Appendix Appendix Specifications Ranges: 0–200 PPM, 0–1999 PPM O , and 0– 2 25% calibration range Sensitivity: 1% of full scale Signal Output: Standard: 0–10 VDC, negative ground 4-20 mA, negative ground plus contact closures for range identification. Display: Digital 7 segment, 3½-digit LCD with over range indicator. Alarms: One high alarm, adjustable & fail-safe. One low alarm, adjustable & fail-safe. One low alarm, fixed at 2% full scale and fail-safe. Accuracy: ±2% of full scale at cons

Summary of the content on the page No. 18

Appendix 319R Spare Parts List QTY. P/N DESCRIPTION 1 C57304 PC board, main 1 A33748 Thermistor assembly 1 C6689-Z2 Micro-Fuel Cell, Z-2 4 F1189 Fuse, 1A (AC) 4 F51 Fuse, ½A (DC) 1 A59264 Probe to analyzer cable, 6 feet A minimum charge of US $20.00 is applicable to spare parts orders. IMPORTANT: Orders for replacement parts should include the part num- ber and the model and serial number of the system for which the parts are intended. Send orders to: TELEDYNE BROWN ENGINEERING Analytical Instru

Summary of the content on the page No. 19

319R Appendix Drawing List C-57490 Outline diagram A-58154 Probe interconnection diagram D-57302 Electronic PCB schematic diagram C-57305 Power supply PCB schematic diagram C-57307 Power supply PCB assembly diagram B-58123 Analysis unit (probe) outline Note: 319R analyzers with 25% and 100% range modifications incorporate the following changes: 100% range modified circuit changes: 1. JP2 - connects pins 1 and 2 2. R52 is 12.5K 3. R30 is 7.5K 4. R35 is 1.5K 25% range modified circuit changes: 1.

Summary of the content on the page No. 20

Appendix 319R Material Safety Data Sheet Section I – Product Identification Product Name: Micro-Fuel Cells and Super Cells, all classes except A-2C, A-3, and A-5. Electrochemical Oxygen Sensors, all classes except R-19. Mini-Micro-Fuel Cells, all classes. Manufacturer: Teledyne Brown Engineering Analytical Instruments Address: 16830 Chestnut Street, City of Industry, CA 91749 Phone: (818) 961-9221 Customer Service: Extension 222 Environmental Health and Safety: Extension 230 Date Prepared : 04/2


Alternative user manuals
# User manual Category Download
1 Teledyne OT-3 User manual Oxygen Equipment 2
2 Teledyne 3000PA User manual Oxygen Equipment 3
3 Teledyne 3000ZA-3X User manual Oxygen Equipment 2
4 Teledyne 3000TA User manual Oxygen Equipment 24
5 Teledyne 3000ZA User manual Oxygen Equipment 1
6 Teledyne 3000TA-XL-EU User manual Oxygen Equipment 8
7 Teledyne 3010MB User manual Oxygen Equipment 5
8 Teledyne 311PC User manual Oxygen Equipment 6
9 Teledyne 3010TBC User manual Oxygen Equipment 1
10 Teledyne 310 User manual Oxygen Equipment 2
11 Teledyne 307WA User manual Oxygen Equipment 0
12 Teledyne 3110 User manual Oxygen Equipment 75
13 Teledyne 311-D User manual Oxygen Equipment 16
14 Teledyne 315 User manual Oxygen Equipment 2
15 Teledyne 3300PA User manual Oxygen Equipment 3
16 3M 8110S User manual Oxygen Equipment 1
17 3M 7187 User manual Oxygen Equipment 1
18 3M 8210PLUS User manual Oxygen Equipment 1
19 3M 8214 User manual Oxygen Equipment 0
20 3M 8210 User manual Oxygen Equipment 6