Congratulations. You have purchased a LifeSource 
state-of-the-art blood pressure monitor, one of the most
technologically advanced yet easy to use products 
available in the marketplace today. This LifeSource 
monitor is designed to make your daily regimen 
useful, and convenient.
Physicians agree that daily self-monitoring of blood 
pressure is an important way individuals can contribute 
to maintaining their cardiovascular health and preventing
the serious consequences of hypertension left
                    
                    ملخص المحتوى في الصفحة رقم 2 
                    
                        TABLE OF CONTENTS WHAT DISPLAY SYMBOLS MEAN . . . . . . . . . . . . . . . . . . . . 3 MONITOR COMPONENTS . . . . . . . . . . . . . . . . . . . . . . . . . 4 ? HOW 702 WORKS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 BEFORE YOU START . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 SELECTING THE CORRECT CUFF . . . . . . . . . . . . . . . . . . . . . . 6 TAKING YOUR BLOOD PRESSURE . . . . . . . . . . . . . . . . 7 — 10 CHECK/ADJUST EXHAUST VELOCITY . . . . . . . . .
                    
                    ملخص المحتوى في الصفحة رقم 3 
                    
                        WHAT DISPLAY SYMBOLS MEAN ? Display Condition/ Recommended Symbol Cause Action Appears during measure- Remain still and refrain ment; blinks and beeps from talking; measurement ❤ MEASURING when pulse is detected. in progress. Appears on display prior to Press exhaust button to and during measurement. release air. Blinks if there is air in cuff ERROR at power up. Measurement begun, but Increase pressurization by pressure is not sufficient for 30 mmHg to 40 mmHg and reading. repeat measurement. RE
                    
                    ملخص المحتوى في الصفحة رقم 4 
                    
                        MONITOR COMPONENTS ? AIR SOCKET SYSTOLIC  DIASTOLIC (PULSE) PRESSURE AND UA-702 PULSE DISPLAY DIGITAL BLOOD PRESSURE MONITOR MEASURING POWER PRESS EXHAUST RE-MEASURE BATTERY LOW RE-MEASURE CUFF POWER BUTTON AIR CONNECTOR CUFF INFLATOR BULB EXHAUST VALVE  AND ADJUSTMENT SCREW BATTERY COMPARTMENT LifeSource Health Line (Toll-Free): 1-888-726-9966 4 ?                                                                                                                                                      
                    
                    ملخص المحتوى في الصفحة رقم 5 
                    
                        HOW 702 WORKS ? LifeSource Blood Pressure Monitors are easy to use, accurate  and digitally display full measurement readouts. Our technology is based on the “oscillometric method” — a noninvasive blood  pressure determination. The term “oscillation” refers to any  measure of vibrations caused by the arterial pulse.  The cuff is first inflated until the artery is fully occluded. Then, the monitor takes measurements while the cuff deflates. Our monitors exam- ine the pulsatile pressure generated 
                    
                    ملخص المحتوى في الصفحة رقم 6 
                    
                        SELECTING THE CORRECT CUFF Using the correct cuff size is important for an accurate reading.  A cuff that is too large will produce a reading that is lower than the correct blood pressure; a cuff that is too small yields a  measurement that is higher. With your arm hanging at the side  of your body, measure the circumference of your upper arm  at the midpoint between shoulder and elbow. RECOMMENDED REPLACEMENT ARM SIZE CUFF SIZE CUFF MODEL# 5.1” - 7.9” (13-20cm) Small UA-279 7.5” - 12.2” (19-31c
                    
                    ملخص المحتوى في الصفحة رقم 7 
                    
                        TAKING YOUR BLOOD PRESSURE Tips for Blood Pressure Monitoring: Tips for Blood Pressure Monitoring: ☞ Relax for about 5 to 10 minutes before measurement. ☞ Relax for about 5 to 10 minutes before measurement. ☞ Remove constricting clothing and place cuff on bare arm. ☞ Remove constricting clothing and place cuff on bare arm. ☞ Unless your physician recommends otherwise, use left arm  ☞ Unless your physician recommends otherwise, use left arm  to measure pressure. to measure pressure. Now you are r
                    
                    ملخص المحتوى في الصفحة رقم 8 
                    
                        TAKING YOUR BLOOD PRESSURE ☞ IMPORTANT: Measure pressure at the same time each day. 4. Press the POWER button. 5. Watch for all display symbols to appear briefly, followed by a  zero, indicating that monitor is ready for measurement. 6. Watch the display and squeeze the bulb quickly until cuff  pressure reaches about 30 mmHg to 40 mmHg above expected  systolic pressure. To help you determine the correct inflation  level, the monitor is programmed to beep quickly three times  at each of the four 
                    
                    ملخص المحتوى في الصفحة رقم 9 
                    
                        TAKING YOUR BLOOD PRESSURE 7. Sit quietly during measurement. Pressure values are displayed at left, then right, along with ❤ symbol blinks and beeps with pulse beat. NOTE: If you wish to stop inflation at any time, press the exhaust  valve to release air. 8. Wait for the long beep, indicating that the measurement is  complete. Your systolic pressure will appear on the left and  your diastolic pressure on the right of the display screen,  alternating with pulse reading.  9. Press the exhaust val
                    
                    ملخص المحتوى في الصفحة رقم 10 
                    
                        TAKING YOUR BLOOD PRESSURE 10.Remove cuff and make a note of your blood pressure  and pulse rate on the chart (see page 16), indicating  date and time of measurement.  We advise that you record the date and time after each measure- ment because an accurate blood pressure history relies not on  single or sporadic readings but on a pattern over time. NOTE:  If you wish to take your blood pressure again, relax and  wait 5–10 minutes to enable the flow of blood in the arm to return  to normal. There
                    
                    ملخص المحتوى في الصفحة رقم 11 
                    
                        CHECK/ADJUST EXHAUST VELOCITY Exhaust velocity is the rate at which the air pressure in the cuff deflates during the course of measurement. It is preset at the  factory and generally does not need any further adjustment. NOTE: The exhaust rate of your monitor was preset and tested at the factory to ensure proper measurement. The rate set is dependent on cuff size, not arm size. If using a cuff other than the one shipped with the unit, you may need to adjust the exhaust velocity rate.  To check t
                    
                    ملخص المحتوى في الصفحة رقم 12 
                    
                        CHECK/ADJUST EXHAUST VELOCITY 4. Watch the display screen. As pressure automatically decreases, diastolic numbers should decline in increments of 2 to 5 mmHg.  If the exhaust velocity range is not within 2 to 5 mmHg, make the following adjustment with the unit turned off: 1. The adjustment screw is located on the exhaust valve  attached to the inflator bulb. It is covered by a thin grey protection cap with a hole in the center. To remove and  discard the cap (you do not need to replace it), inse
                    
                    ملخص المحتوى في الصفحة رقم 13 
                    
                        ABOUT BLOOD PRESSURE ❤ What Is Blood Pressure? What Is Blood Pressure? Blood pressure is the force exerted by blood against the Blood pressure is the force exerted by blood against the walls of  walls of the arteries. Systolic pressure occurs when the heart the arteries. Systolic pressure occurs when the heart contracts;  contracts; diastolic pressure occurs when the heart expands. diastolic pressure occurs when the heart expands. Blood pressure  Blood pressure is measured in millimeters of merc
                    
                    ملخص المحتوى في الصفحة رقم 14 
                    
                        ABOUT BLOOD PRESSURE ❤ Assessing High Blood Pressure Assessing High Blood Pressure The following standards for assessing high blood pressure The following standards for assessing high blood pressure  (without regard to age) have been established as a guideline. (without regard to age) have been established by the National Institutes of Health JNCVI. BP Classification Systolic (mmHg) Diastolic (mmHg) Category Systolic (mmHg) Diastolic (mmHg) Normal <120 and <80 Optimal <120 <80 Prehypertension 12
                    
                    ملخص المحتوى في الصفحة رقم 15 
                    
                        ABOUT BLOOD PRESSURE ❤ Why Measure Blood Pressure at Home? Why Measure Blood Pressure at Home? It is now well known that, for many individuals, blood pres- It is now well known that, for many individuals, blood  sure readings taken in a doctor’s office or hospital setting pressure readings taken in a doctor’s office or hospital setting might be elevated as a result of apprehension and anxiety. might be elevated as a result of apprehension and anxiety.  This response is commonly called “white coa
                    
                    ملخص المحتوى في الصفحة رقم 16 
                    
                        TM BLOOD PRESSURE RECORD Name:                                                       Age:               Weight:  DATE          AM         SYS/DIA       PULSE        PM       SYS/DIA       PULSE 1/14 9:30 132/98 P 63 P 69 6:30 128/87 EXAMPLE LifeSource Health Line (Toll-Free): 1-888-726-9966 16 ❤                                                                                                                                                                                                            
                    
                    ملخص المحتوى في الصفحة رقم 17 
                    
                        IMPORTANT INFORMATION ✓ Please read this important information before using your monitor. ☞ Please remember that only a medical practitioner is qualified to interpret your blood pressure measurements. Use of this  device should not replace regular medical examinations. ☞ Have your physician review your procedure for using this  device. He or she will want to verify blood pressure  measurements before prescribing or adjusting medications.  ☞ Consult your physician if you have any doubt about your
                    
                    ملخص المحتوى في الصفحة رقم 18 
                    
                        SPECIFICATIONS SPECIFICATIONS ✓ Model . . . . . . . . . . . . . . . . . . .UA-702 Type . . . . . . . . . . . . . . . . . . . . .Oscillometric Display . . . . . . . . . . . . . . . . . .Digital, 10-mm character height Pressure/pulse displayed alternately Measurement range . . . . . . . . .Pressure: 20 mmHg to 280 mmHg Pulse: 40 pulses to 200  pulses/minute Accuracy . . . . . . . . . . . . . . . . .Pressure: +3 mmHg or 2%,  whichever is greater Pulse: +5% Pressurization . . . . . . . . . . . .Manu
                    
                    ملخص المحتوى في الصفحة رقم 19 
                    
                        CONTACT INFORMATION Our products are designed and manufactured using the latest  Our products are designed and manufactured using the latest scientific and technological methods, and offer accurate, easy to scientific and technological methods, and offer accurate, easy use, home monitoring and treatment options. The product  to use, home monitoring and treatment options. Our full product line includes: line includes: Aneroid Blood Pressure Kits Aneroid Blood Pressure Kits Blood Pressure Cuffs Bl
                    
                    ملخص المحتوى في الصفحة رقم 20 
                    
                        CONTACT INFORMATION IMPORTANT! If You Need Assistance with the Set-Up or Operation We Can Help! Please call us FIRST before contacting your retailer at LifeSource Health Line 1-888-726-9966 - Toll Free 1-800-463-5414 - In Canada A specially trained representative will assist you A division of A&D Engineering, Inc.  1555 McCandless Drive  Milpitas, CA 95035 www.LifeSourceOnline.com LifeSource is a trademark of A&D Medical.  ©2003 A&D Medical. All rights reserved.  Specifications subject to change