ملخص المحتوى في الصفحة رقم 1 
                    
                        Owner’s Operator and Maintenance Manual
MA90 Series
MA90Z Alternating Rotation 
with On Demand True Low 
Air Loss System
MA95Z Alternating Rotation 
and True Low Air Loss System
DEALER: This manual MUST be given to 
the user of the product.
USER: BEFORE using this product, read this 
manual and save for future reference.
For more information regarding 
Invacare products,  parts, and services, 
please visit www.invacare.com                                                                          
                    
                    ملخص المحتوى في الصفحة رقم 2 
                    
                        SPECIAL NOTES  WARNING DO NOT use this product or any available optional equipment without first  completely reading and understanding these instructions and any additional  instructional material such as owner’s manuals, service manuals or instruction  sheets supplied with this product or optional equipment. If you are unable to  understand the warnings, cautions or instructions, contact a healthcare  professional, dealer or technical personnel before attempting to use this equipment  - otherw
                    
                    ملخص المحتوى في الصفحة رقم 3 
                    
                        TABLE OF CONTENTS TABLE OF CONTENTS SPECIAL NOTES ................................................................................ 2 REGISTER YOUR PRODUCT ...............................................................4 TYPICAL PRODUCT PARAMETERS .................................................... 5 Electrical Parameters.............................................................................................................................5 Performance Parameters............................
                    
                    ملخص المحتوى في الصفحة رقم 4 
                    
                        TABLE OF CONTENTS TABLE OF CONTENTS Low Pressure LED...............................................................................................................................19 Displaying the Total Run Time (MA95Z).......................................................................................19 Powering Up the System ........................................................................................................................19 Placing the Patient on the Mattress..........
                    
                    ملخص المحتوى في الصفحة رقم 5 
                    
                        TYPICAL PRODUCT PARAMETERS TYPICAL PRODUCT PARAMETERS Electrical Parameters MA90Z MA95Z INPUT VOLTAGE AC: 90 V INPUT FREQUENCY: 60 Hz CURRENT: 1 A 2 A MAXIMUM POWER CONSUMPTION: 30 ± 10 W 150 ± 10 W CIRCUIT PROTECTION: Dual fused, Dual fused,  250 V, 250 V,  1 A fast blow fuses 5 A fast blow fuses MODE OF OPERATION: Continuous Performance Parameters MA90Z MA95Z WEIGHT CAPACITY STANDARD MATTRESS: 350 lbs BARIATRIC MATTRESS: 1000 lbs PRESSURE ZONE: 2 MAXIMUM FLOW: 50 ± 15 LPM 1275 ± 100 LPM MAXIMU
                    
                    ملخص المحتوى في الصفحة رقم 6 
                    
                        TYPICAL PRODUCT PARAMETERS Mechanical Parameters MA90Z MA95Z CONTROL UNIT DIMENSIONS (L X W X H): 15” x 6½” x 11” CONTROL UNIT WEIGHT: 15 lbs CONTROL UNIT POWER CORD: 10 - 14 Feet Long, Hospital Grade CONTROL UNIT CONNECTION: Quad Flow - Four ¼” Triple Flow - Two ½” and  Single Quick Coupling One ¼” Single Magnetic  Quick Coupling CONTROL UNIT PACKAGING: 1 piece/box AIR FILTER: None Charcoal Air Filter with  Fire Retardant Environmental Parameters MA90Z/MA95Z OPERATING CONDITIONS AMBIENT TEMPERA
                    
                    ملخص المحتوى في الصفحة رقم 7 
                    
                        LABEL LOCATION LABEL LOCATION DANGER-EXPLOSION HAZARD: DO NOT use in the presence of  flammable anesthetics. CAUTION: Equipment should be connected to a properly grounded  receptacle (3-prong). Risk of Electrical shock. DO NOT remove  back. Disconnect air hose before administering CPR. NOTE: This label is on the back of  the unit and also contains the   ATTENTION serial number. Clean Filter Every 5 months and  Whenever Dirty. Remove Filter,  Wash, Dry & Replace. NOTE: MA95Z only.  WARNING Pati
                    
                    ملخص المحتوى في الصفحة رقم 8 
                    
                        SECTION 1—GENERAL GUIDELINES SECTION 1—GENERAL GUIDELINES  WARNING SECTION 1 - GENERAL GUIDELINES contains important information for the safe  operation and use of this product. Contraindications ALWAYS consult the patient’s physician before using the MA90Z and MA95Z systems. Installation The MA90Z and MA95Z systems MUST be installed on medical bed frames with side  rails. The side rails MUST be in the raised position whenever a patient is on the bed. Controls on the footboard may be obstructed
                    
                    ملخص المحتوى في الصفحة رقم 9 
                    
                        SECTION 1—GENERAL GUIDELINES Fire Hazard  DANGER Smoking DO NOT SMOKE while using this device. This system uses room air for circulation  through the mattress. A cigarette can burn a hole in the bed surface and cause  damage to the mattress. Also, patient clothing, bed sheets, etc. may be combustible  and cause a fire. Failure to observe this warning can result in severe fire, property  damage and cause physical injury or death. Smoking by visitors in the room will contaminate the system. There
                    
                    ملخص المحتوى في الصفحة رقم 10 
                    
                        SECTION 2—OPERATION SECTION 2—OPERATION CAUTION The control unit and mattress on the MA90Z series are designed to be used as a  system. DO NOT replace mattresses or control units with other models or other  brands. Otherwise, damage to the system may occur. Contact your supplier to get  the correct replacement if needed. Installing the Mattress Replacement System CAUTION DO NOT strap the mattress to the bed frame at the head and foot ends. Secure  ALL mattress straps. Secure the straps to the be
                    
                    ملخص المحتوى في الصفحة رقم 11 
                    
                        SECTION 2—OPERATION Head End Powered Mattress NOTE: Mattress  cover not shown for  clarity. Buckle  Straps Hose Foot End FIGURE 2.1   Installing the Mattress Replacement System Installing the Side Rails  WARNING Patient entrapment with bed side rails may cause injury or death. Mattress MUST  fit bed frame and side rails snugly to prevent patient entrapment. Follow the  manufacturer’s instructions. Monitor patient frequently. Read and understand the  Owner’s/Operator’s Manual prior to using this
                    
                    ملخص المحتوى في الصفحة رقم 12 
                    
                        SECTION 2—OPERATION Connecting the Hose CAUTION Ensure that the hose connecting the control unit to the mattress is routed such that  it cannot be stepped on, kinked, squeezed or otherwise damaged. NOTE: For this procedure, refer to FIGURE2.3. 1. Locate the hose at the foot end of the mattress. 2. Locate the control unit connector on the right side of the control unit. 3. Perform one of the following: • MA90Z ‐ Perform the following steps (Detail “A”): i. Squeeze and hold the tabs on the hose co
                    
                    ملخص المحتوى في الصفحة رقم 13 
                    
                        SECTION 2—OPERATION Control  Unit DETAIL “B” - MA95Z  CONNECTOR Control Unit  Connectors Connector  Cover DETAIL “A” - MA90Z  CONNECTOR MA95Z Hose  Connector MA90Z Hose  Tab Connector FIGURE 2.3   Connecting the Hose Connecting the Power Cord  WARNING DO NOT alter plug to fit a non-conforming outlet. Instead, have an electrician  install a properly grounded outlet. Failure to use the correct plug and outlet can  result in a potential safety hazard. CAUTION Ensure that the power cord of the cont
                    
                    ملخص المحتوى في الصفحة رقم 14 
                    
                        SECTION 2—OPERATION Using the Front Panel NOTE: For this procedure, refer to FIGURE2.4. DETAIL “A” - MA90Z FRONT PANEL Low Pressure  Turn Angle  Power LED Power  Button Button Fail LED Lock/Alarm  Turn Button Mode (Max  Button Soft Button Inflate/Low Air  Firm Button Turn Time Button Loss) Button DETAIL “B” - MA95Z FRONT PANEL Low Pressure  Hold  Power  Power  LED Select Button Button Fail LED Button Lock/Alarm  Rotation  Max Inflate  Button Button Down/Soft  LCD Up/Firm Button Button Button FIG
                    
                    ملخص المحتوى في الصفحة رقم 15 
                    
                        SECTION 2—OPERATION Turn Button (MA90Z) 1. Press the Turn button () to select one of three rotation modes or to set the unit to  No Turn (Static) mode.  NOTE: The rotation modes are right, left or both and all three modes include alternating pressure  therapy. NOTE: When No Turn is selected, all of the air cushions are maintained at a constant pressure. Turn Time Button (MA90Z) 1. Press the Turn Time button () to select one of four rotation times.  NOTE: The rotation times are 10, 20, 30 and 60 
                    
                    ملخص المحتوى في الصفحة رقم 16 
                    
                        SECTION 2—OPERATION FUNCTION DESCRIPTION PULSE Pulse mode. The pulse time indicates the time between  pulses. For example, 5 minutes is selected. In this  example, all of the air cushions in the mattress will  deflate for 4 seconds  and inflate for 5 minutes (the  selected time). FOWLER Fowler Mode. The pressure will increase to prevent the  patient from bottoming out. MULTI (PULSE + STATIC) Pulse and Static mode.  3. Use the Select button to move the pointer to the times or attributes available
                    
                    ملخص المحتوى في الصفحة رقم 17 
                    
                        SECTION 2—OPERATION Programming the Automatic Wireless Fowler Transmitter 1. Ensure the transmitter is inside the mattress pocket located on the right side of the  mattress at the head end. 2. Ensure the mattress is in a flat position. 3. Ensure the control unit is plugged in, and the LCD displays STAND BY. 4. Perform one of the following: • MA90Z ‐ Perform the following steps: i. Press and hold the Turn Angle and Max Inflate buttons. NOTE: The Fowler LED is lit and L displays in the LCD Display
                    
                    ملخص المحتوى في الصفحة رقم 18 
                    
                        SECTION 2—OPERATION Mode Button (MA90Z) 1. Press the Mode button () to select the Max Inflate Mode or the Low Air Loss mode. •Max Inflate Mode ‐ In this mode, the mattress inflates rapidly to maximum  firmness (pressurized to 35 ± 6 mmHg). A series of beeps sound every three  minutes as a reminder that the Max Inflate mode is active. After 30 minutes, the  Max Inflate mode deactivates and the control unit defaults to the previous setting.  NOTE: It takes 5‐10 minutes for the MA90Z mattress to in
                    
                    ملخص المحتوى في الصفحة رقم 19 
                    
                        SECTION 2—OPERATION Lock/Alarm Silence Button 1. Press the Lock/Alarm Silence button () to select one of the following modes:  •Lock Mode ‐ Pressing this button locks out all control unit functions, including the  On button, to prevent any tampering with the settings. Press the button for  approximately 3‐5 seconds for the Lock LED to activate. NOTE: The Lock LED illuminates when in this mode.  •Alarm Silence mode ‐ Pressing this button silences the alarm that sounds in the  event of power failu
                    
                    ملخص المحتوى في الصفحة رقم 20 
                    
                        SECTION 2—OPERATION Placing the Patient on the Mattress 1. Press the Mode button or Max Inflate button to turn on the control unit to maximum  flow.  NOTE: In this mode, the Max Inflate LED lights up. 2. After the mattress is fully inflated, place the patient on the bed. 3. Ensure that the patient’s feet are towards the foot end of the mattress (the end with the  hose).  4. Center the patient on the bed from side‐to‐side and head‐to‐foot. NOTE: Special positioning may be required with contracted