Résumé du contenu de la page N° 1 
                    
                        Owner’s Operator and Maintenance Manual
MA80 Series
MA80 True Low Air Loss 
System
MA85 Five-Zone Alternating 
Pressure and True Low Air 
Loss System
DEALER: This manual MUST be given to 
the user of the product.
USER: BEFORE using this product, read this 
manual and save for future reference.
For more information regarding 
Invacare products,  parts, and services, 
please visit www.invacare.com                                                                                                      
                    
                    Résumé du contenu de la page N° 2 
                    
                        SPECIAL NOTES  WARNING DO NOT use this product or any available optional equipment without first  completely reading and understanding these instructions and any additional  instructional material such as owner’s manuals, service manuals or instruction  sheets supplied with this product or optional equipment. If you are unable to  understand the warnings, cautions or instructions, contact a healthcare  professional, dealer or technical personnel before attempting to use this equipment  - otherw
                    
                    Résumé du contenu de la page N° 3 
                    
                        TABLE OF CONTENTS TABLE OF CONTENTS SPECIAL NOTES ................................................................................2 REGISTER YOUR PRODUCT ............................................................... 4 TYPICAL PRODUCT PARAMETERS .................................................... 5 Electrical Parameters.............................................................................................................................5 Performance Parameters............................
                    
                    Résumé du contenu de la page N° 4 
                    
                        TABLE OF CONTENTS TABLE OF CONTENTS Transferring Patient From/To a Gurney .............................................................................................19 Transferring Patient From/To a Wheelchair......................................................................................19 Preparing for CPR Procedure................................................................................................................19 SECTION 3—MAINTENANCE AND TROUBLESHOOTING ................
                    
                    Résumé du contenu de la page N° 5 
                    
                        TYPICAL PRODUCT PARAMETERS TYPICAL PRODUCT PARAMETERS Electrical Parameters MA80/MA85 INPUT VOLTAGE AC: 90 V INPUT FREQUENCY: 60 Hz CURRENT: 2 A MAXIMUM POWER CONSUMPTION: 150 ± 10 W CIRCUIT PROTECTION: Dual fused,  250 V,  5 A fast blow fuses MODE OF OPERATION: Continuous Performance Parameters MA80 MA85 WEIGHT CAPACITY STANDARD MATTRESS: 350 lbs BARIATRIC MATTRESS: 1000 lbs PRESSURE ZONE: 5 MAXIMUM FLOW: 1275 ± 100 LPM MAXIMUM PRESSURE: 35 ± 5 mmHg MAXIMUM FLOW TIMER: 15 minutes SUPPORT SURFAC
                    
                    Résumé du contenu de la page N° 6 
                    
                        TYPICAL PRODUCT PARAMETERS Environmental Parameters MA80/MA85 OPERATING CONDITIONS AMBIENT TEMPERATURE: 50° - 104° F RELATIVE HUMIDITY: 30% - 75% Non-Condensing ATMOSPHERIC PRESSURE: 70 - 106 kPa STORAGE AND SHIPPING CONDITIONS AMBIENT TEMPERATURE: -40° - 158° F RELATIVE HUMIDITY: 10% - 100%  ATMOSPHERIC PRESSURE: 50 - 106 kPa MA80 Series 6 Part No 1148138                                                                                                                                              
                    
                    Résumé du contenu de la page N° 7 
                    
                        LABEL LOCATION LABEL LOCATION DANGER-EXPLOSION HAZARD: DO NOT use in the presence of  flammable anesthetics. CAUTION: Equipment should be connected to a properly grounded  receptacle (3-prong). Risk of Electrical shock. DO NOT remove  back. Disconnect air hose before administering CPR. NOTE: This label is on the back of  the unit and also contains the   ATTENTION serial number. Clean Filter Every 5 months and  Whenever Dirty. Remove Filter,  Wash, Dry & Replace.  WARNING Patient entrapment wit
                    
                    Résumé du contenu de la page N° 8 
                    
                        SECTION 1—GENERAL GUIDELINES SECTION 1—GENERAL GUIDELINES  WARNING SECTION 1 - GENERAL GUIDELINES contains important information for the safe  operation and use of this product. Contraindications ALWAYS consult the patient’s physician before using the MA80 and MA85 systems. Installation The MA80 and MA85 systems MUST be installed on medical bed frames with side rails.  The side rails MUST be in the raised position whenever a patient is on the bed. Controls on the footboard may be obstructed by 
                    
                    Résumé du contenu de la page N° 9 
                    
                        SECTION 1—GENERAL GUIDELINES Fire Hazard  DANGER Smoking DO NOT SMOKE while using this device. This system uses room air for circulation  through the mattress. A cigarette can burn a hole in the bed surface and cause  damage to the mattress. Also, patient clothing, bed sheets, etc. may be combustible  and cause a fire. Failure to observe this warning can result in severe fire, property  damage and cause physical injury or death. Smoking by visitors in the room will contaminate the system. There
                    
                    Résumé du contenu de la page N° 10 
                    
                        SECTION 2—OPERATION SECTION 2—OPERATION CAUTION The control unit and mattress on the MA80 series are designed to be used as a  system. DO NOT replace mattresses or control units with other models or other  brands. Otherwise, damage to the system may occur. Contact your supplier to get  the correct replacement if needed. Installing the Mattress Replacement System CAUTION DO NOT strap the mattress to the bed frame at the head and foot ends. Secure  ALL mattress straps. Secure the straps to the bed
                    
                    Résumé du contenu de la page N° 11 
                    
                        SECTION 2—OPERATION Head End Powered Mattress Buckle  Straps Hose Foot End FIGURE 2.1   Installing the Mattress Replacement System Installing the Side Rails  WARNING Patient entrapment with bed side rails may cause injury or death. Mattress MUST  fit bed frame and side rails snugly to prevent patient entrapment. Follow the  manufacturer’s instructions. Monitor patient frequently. Read and understand the  Owner’s/Operator’s Manual prior to using this equipment. Invacare product  manuals are avai
                    
                    Résumé du contenu de la page N° 12 
                    
                        SECTION 2—OPERATION Connecting the Hose CAUTION Ensure that the hose connecting the control unit to the mattress is routed such that  it cannot be stepped on, kinked, squeezed or otherwise damaged. NOTE: For this procedure, refer to FIGURE 2.3. 1. Locate the hose at the foot end of the mattress. 2. Locate the control unit connector on the right side of the control unit. 3. Lift and hold the connector cover up. 4. Push the hose connector (Detail “A”) onto the control connector. 5. Ensure the hose
                    
                    Résumé du contenu de la page N° 13 
                    
                        SECTION 2—OPERATION Connecting the Power Cord  WARNING DO NOT alter plug to fit a non-conforming outlet. Instead, have an electrician  install a properly grounded outlet. Failure to use the correct plug and outlet can  result in a potential safety hazard. CAUTION Ensure that the power cord of the control unit is not pinched, or has any objects  placed on it, and also ensure it is not located where it can be stepped on or tripped  over. NOTE: For this procedure, refer to FIGURE 2.4. 1. Examine t
                    
                    Résumé du contenu de la page N° 14 
                    
                        SECTION 2—OPERATION Using the Front Panel NOTE: For this procedure, refer to FIGURE 2.5. DETAIL “A” - MA80 FRONT PANEL Low Pressure  Fowler  Power LED Power Fail  Button Pulse Button Button LED Lock/Alarm  Static Button Max Inflate  Soft Button Button LCD  Button Firm Button Display DETAIL “B” - MA85 FRONT PANEL Low Pressure  Fowler  Power Fail  Power  Alternating  LED Button LED Button Pressure Button Lock/Alarm  Static Button Max Inflate  Button Soft Button LCD  Firm Button Button Display FIGU
                    
                    Résumé du contenu de la page N° 15 
                    
                        SECTION 2—OPERATION Pulse Button (MA80) 1. Press the Pulse button ( ) to Turn on Pulse mode and select one of four Pulse Times  - 1, 3, 5 or 10 minutes.  NOTE: The pulse times are 1, 3, 5 or 10 minutes. The pulse times indicate the time between each  low pulse. For example, if 5 minutes is selected, all the air cushions in the mattress will inflate for  5 minutes (the selected time) to the selected comfort setting and then drop 20% pressure for 20  seconds. Alternating Pressure Button (MA85) 1. 
                    
                    Résumé du contenu de la page N° 16 
                    
                        SECTION 2—OPERATION Static Button 1. Press the Static button ( ) to select the static mode and maintain the mattress at  constant pressure. NOTE: The Static LED is illuminated GREEN. Firm/Soft Buttons 1. Select comfort pressure settings by pressing Firm ( ) or Soft ( ) buttons. • Soft Button - Pressing this button reduces the pressure in the mattress. • Firm Button - Pressing this button increases the pressure in the mattress. NOTE: The patient comfort pressure ranges from Soft (level 0 = 6±4 mm
                    
                    Résumé du contenu de la page N° 17 
                    
                        SECTION 2—OPERATION 5. Perform one of the following: • Cancel - Press the Fowler button to exit Fowler Transmitter Setup without linking  to a transmitter. • Save - Wait 60 seconds without pressing a key.  Max Inflate Button 1. Press the Max Inflate button ( ) to select the max inflate mode and inflate the  mattress rapidly to maximum firmness (pressurized to 35 ± 6 mmHg). A series of  beeps sound every three minutes as a reminder that the Max Inflate mode is active.  After 30 minutes, the Max I
                    
                    Résumé du contenu de la page N° 18 
                    
                        SECTION 2—OPERATION Powering Up the System NOTE: For this procedure, refer to FIGURE 2.5 on page 14. 1. Turn on the power to the system by pressing the Power button on the control unit. NOTE: Once the button is released, a GREEN LED illuminates when the unit is on. Placing the Patient on the Mattress 1. Press the Max Inflate button to turn on the control unit to maximum flow.  NOTE: In this mode, the Max Inflate LED lights up. 2. After the mattress is fully inflated, place the patient on the bed
                    
                    Résumé du contenu de la page N° 19 
                    
                        SECTION 2—OPERATION Transferring Patient From/To a Gurney  WARNING ALWAYS engage the wheel locks of the bed and the wheel locks of the gurney  before transferring the patient between the bed and the gurney. 1. Engage the wheel locks of the bed. Refer to the owner’s manual provided with the bed. 2. Engage the wheel locks of the gurney. Refer to the owner’s manual provided with the  gurney. 3. Press the Max Inflate button to achieve maximum mattress pressure. 4. Raise or lower the bed to match th
                    
                    Résumé du contenu de la page N° 20 
                    
                        SECTION 2—OPERATION DETAIL “A” Connector  Cover Control Unit  Connector Control  Unit DETAIL “B” Hose  Connector Holes FIGURE 2.6   Preparing for CPR Procedure MA80 Series 20 Part No 1148138