ページ1に含まれる内容の要旨 
                    
                        *TB 9-6625-1996-35 
DEPARTMENT OF THE ARMY TECHNICAL BULLETIN 
CALIBRATION PROCEDURE FOR 
FREQUENCY COUNTER 
HEWLETT-PACKARD 
MODELS 5345A, 5345A/E28 AND 5345AOPT12 
Headquarters, Department of the Army, Washington, DC 
25 June 2001 
Approved for public release; distribution is unlimited 
REPORTING OF ERRORS AND RECOMMENDED IMPROVEMENTS 
You can help improve this publication.  If you find any mistakes or if you know of a 
way to improve the procedure, please let us know.  Mail your letter or DA 
                    
                    ページ2に含まれる内容の要旨 
                    
                        TB 9-6625-1996-35  SECTION I  IDENTIFICATION AND DESCRIPTION  1. Test Instrument Identification.  This bulletin provides instructions for the  calibration of Frequency Counter, Hewlett-Packard, Models 5345A, 5345A/E28 and  5345AOPT12.  The manufacturer's manuals were used as the prime data sources in  compiling these instructions.  The equipment being calibrated will be referred to as the  TI (test instrument) throughout this bulletin.   a. Model Variations.  Differences among models are listed 
                    
                    ページ3に含まれる内容の要旨 
                    
                        TB 9-6625-1996-35  SECTION II  EQUIPMENT REQUIREMENTS  4. Equipment Required.  Table 2 identifies the specific equipment to be used in this  calibration procedure.  This equipment is issued with Secondary Transfer Calibration  Standards Set AN/GSM-287 or AN/GSM-705.  Alternate items may be used by the  calibrating activity.  The items selected must be verified to perform satisfactorily prior  to use and must bear evidence of current calibration.  The equipment must meet or  exceed the minimum us
                    
                    ページ4に含まれる内容の要旨 
                    
                        TB 9-6625-1996-35  SECTION III  CALIBRATION PROCESS  6. Preliminary Instructions   a. The instruction outlined in paragraphs 6 and 7 are preparatory to the calibration  process.  Personnel should become familiar with the entire bulletin before beginning the  calibration.   b. Items of equipment used in this procedure are referenced within the text by  common name as listed in table 2.   c. Unless otherwise specified, verify the result of each test and, whenever the test  requirement is not met, 
                    
                    ページ5に含まれる内容の要旨 
                    
                        TB 9-6625-1996-35   (4) SAMPLE RATE control fully ccw.   (5) CHANNEL A and B LEVEL controls to PRESET.   (6) CHANNEL A and B SLOPE switches to + (positive).   (7) CHANNEL A and B input impedance switches to 50 W.   (8) CHANNEL A and B ATTEN switches to X1.   (9) CHANNEL A and B AC-DC switches to DC.   (10) CHECK-COM A-SEP switch to SEP.  8. Time Base Stability   a. Performance Check    (1) Connect time/frequency workstation OUTPUT 1 MHz to frequency difference  meter REF INPUT.   (2) Connect TI 
                    
                    ページ6に含まれる内容の要旨 
                    
                        TB 9-6625-1996-35  9. Sensitivity   a. Performance Check   (1) Set function generator as listed in (a) through (c) below:   (a) Output Type to UnBalanced 600   (b) Function to Sine Wave.   (b) Frequency to 10 Hz.   (c) Amplitude to 1 mV.   (2) Position TI controls as listed in (a) through (f) below:   (a) CHANNEL A LEVEL control to PRESET.   (b) CHANNEL A SLOPE switch to +.   (c) CHANNEL A 50W/1MW switch to 50W.   (d) CHANNEL A ATTEN X1/X10 switch to X1.   (e) CHANNEL A AC/DC switch to DC.   (f)
                    
                    ページ7に含まれる内容の要旨 
                    
                        TB 9-6625-1996-35   b. Adjustments   (1) Substitute signal generator for function generator.   (2) Connect signal generator RF OUTPUT connector to TI CHANNEL A input.   (3) Connect signal generator EXT REF INPUT connector (rear panel) to TI  FREQ STD OUTPUT 10 MHz connector (rear panel).   (4)  Extend front panel display assembly using extender cable 05345-60205.   Disconnect CHAN A P1 caple from A9 J1 CHA (CHAN B P2 cable to A9 J2 CH B for  CHANNEL B) and connect A9 J1 CH A (A9 J2 CH B for CHAN
                    
                    ページ8に含まれる内容の要旨 
                    
                        TB 9-6625-1996-35   (10) Repeat (8) and (9) above as required to compensate for interaction.   (11) Disconnect oscilloscope and signal generator from TI, reinstall TI from panel,  and reconnect CHAN A P1 cable to A9 J1 CH A and CHAN B P2 cable to A9 J2 CH B.   (12) Replace TI covers.  10. A12 Interface I/O   a. Performance Check  NOTE  This option can only be calibrated in an automated mode.   The TI must be connected to a controller with a properly  configured GPIB IEEE interface card.  NOTE  I
                    
                    ページ9に含まれる内容の要旨 
                    
                        TB 9-6625-1996-35  A:00I1 CHAN A TRIG LEVEL 1.9995 2.0005 b2  B000I1 CHAN B TRIG LEVEL -2.0005 -1.9995 b3  B:00I1 CHAN B TRIG LEVEL 1.9995 2.0005 b4     b. Adjustments  NOTE  Remove top cover to make adjustments, then replace top  cover before repeating (a) above.   (1) Adjust A12 pot A (R69) (fig. 3) for a –2.0000 ± 0.0005 V dc multimeter  indication (R).   (2) Adjust A12 pot B (R74) (fig. 3) for a 2.0000 ± 0.0005 V dc multimeter indication  (R).   (3) Adjust A12 pot D (R86) (fig. 3) for a –2.0
                    
                    ページ10に含まれる内容の要旨 
                    
                        TB 9-6625-1996-35   a. Performance Check   (1) Mount circuit board A6 (fig. 1) on extender boards.   (2) Connect multimeter between A6TP6 and + 5 V RET (fig. 4).  If multimeter  does not indicate between 4.98 and 5.02 V dc, perform b(1) below.   (3) Connect multimeter between A6TP3 and - 5.2 V RET (fig. 4).  If multimeter  does not indicate between -5.18 and -5.22 V, perform b(2) below.   (4) Replace circuit board A6 in TI and mount circuit board A7 (fig. 1) on extender  boards.   (5) Connect mu
                    
                    ページ11に含まれる内容の要旨 
                    
                        TB 9-6625-1996-35    R18 R24 +15V RET TP7 TP2 A7 CIRCUIT BOARD S/N 1428A -15V RET   TP3 R19 R29 +5V RET TP6 -5.2 V RET A6 CIRCUIT BOARD   R16 R17 TP3 TP1 -15V RET +15V RET A7 CIRCUIT BOARD S/N  1808   Figure 4.  Power supply circuit boards.  11                                                                                                                                                                                                                                                                
                    
                    ページ12に含まれる内容の要旨 
                    
                        TB 9-6625-1996-35  12. Final Procedure   a. Deenergize and disconnect all equipment.   b. Annotate and affix DA label/form in accordance with TB 750-25. 12                                                                                                                                                                                                                                                                                                                                                         
                    
                    ページ13に含まれる内容の要旨 
                    
                        TB 9-6625-1996-35    By Order of the Secretary of the Army:  011  6212    Distribution:       To be distributed in accordance with IDN 342172, requirements for calibration procedure   TB 9-6625-1996-35.  PIN:  046669-000                                                                                                                                                                                                                                                                                        
                    
                    ページ14に含まれる内容の要旨