INSTRUCTION MANUAL
 
MPR-513
 
 
MPR-513R 
 
 
MPR-1013 
Pharmaceutical Refrigerator 
 
 
MPR-1013R
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
MPR-513R MPR-1013R 
 
 
 
 
 
 
                                                                                                                                                                                                                                                                                                            
                    
                    ページ2に含まれる内容の要旨 
                    
                            Note:    1. No part of this manual may be reproduced in any form without the expressed written  permission of SANYO.    2. The contents of this manual are subject to change without notice.    3. Please contact SANYO if any point in this manual is unclear or if there are any  inaccuracies.    SANYO Electric Biomedical Co., Ltd. All rights reserved. Printed in Japan.          CONTENTS    PRECAUTIONS FOR SAFE OPERATION   P. 2    CAUTIONS FOR USAGE      P. 6    ENVIRONMENTAL CONDITIONS    P. 7  
                    
                    ページ3に含まれる内容の要旨 
                    
                           PRECAUTIONS FOR SAFE OPERATION    It is imperative that the user complies with this manual as it contains  important safety advice.     Items and procedures are described so that you can use this unit correctly and safely.   If the precautions advised are followed, this will prevent possible injury to the user and  any other person.    Precautions are illustrated in the following way:      WARNING     Failure to observe WARNING signs could result in a hazard to personnel  possibly resulting i
                    
                    ページ4に含まれる内容の要旨 
                    
                           PRECAUTIONS FOR SAFE OPERATION        WARNING       Do not use the unit outdoors.  Current leakage or electric shock may result if the unit is exposed to    rain water.    Only qualified engineers or service personnel should install the unit.  The installation by    unqualified personnel may cause electric shock or fire.    Install the unit on a sturdy floor.  If the floor is not strong enough or the installation site is not    adequate, this may result in injury from the unit falling or tipp
                    
                    ページ5に含まれる内容の要旨 
                    
                           PRECAUTIONS FOR SAFE OPERATION        WARNING       Do not pull out more than 2 drawers at the same time when storing or taking out of the stocked    material.  This may cause injury by falling or tipping of the unit if something heavy items are stored.    Disconnect the power supply to the unit prior to any repair or maintenance of the unit in order to    prevent electric shock or injury.    Ensure you do not inhale or consume medication or aerosols from around the unit at the time of    mai
                    
                    ページ6に含まれる内容の要旨 
                    
                           PRECAUTIONS FOR SAFE OPERATION        CAUTION       Fix the shelves securely.  Incomplete installation may cause injury or damage.      When removing the plug from the power supply outlet, grip the power supply plug, not the cord.     Pulling the cord may result in  electric shock or fire by short circuit.    Never damage or break the power supply plug or cord.  Do not use the supply plug if its cord  is loose.  This may cause fire or electric shock.    Do not touch any electrical parts such 
                    
                    ページ7に含まれる内容の要旨 
                    
                           CAUTIONS FOR USAGE    1.  If the unit is unplugged or the power to the unit is interrupted, do not restart the unit for at least 5  minutes.  This protects the compressor.    2.  This inner cabinet is refrigerated by forced circulation of cooled air inside the chamber.  Ensure that  the intake and exhaust air vents are not blocked.    3.  Adequate space should be provided between the items inside the unit to allow air circulation.     4.  The temperature alarm may be operated at the time of f
                    
                    ページ8に含まれる内容の要旨 
                    
                           CAUTIONS FOR USAGE      Circuit breaker    This unit is equipped with a circuit breaker on the back.  Make sure to switch ON this breaker before the  unit starts to run.  Following figures show the circuit breaker position.      Circuit breaker                   When the operation of the unit is stopped by this breaker, contact a dealer or a service station after  disconnected the power supply plug.    WARNING  Never disassemble, repair, or modify the unit yourself.  Any such work carried out
                    
                    ページ9に含まれる内容の要旨 
                    
                           REFRIGERATOR COMPONENTS      14 1 7 2 8 14         10       11   4      5 17         12       6   9      3     MPR-513R        15 16 13 2 8      1        14 14    10    12    11    4      5  10      7 9                3          MPR-1013R    15 6 16 13 8                                                                                                                                                                                                                                                   
                    
                    ページ10に含まれる内容の要旨 
                    
                           REFRIGERATOR COMPONENTS    1. Door:  Sliding type.  The recessed portion on the rail enables the self-closing of the door.  The glass is pair  construction.  2. Control panel:  Panel can be opened when upper right corner of the lower part cover is pushed.  Refer to page 10.  3. Space for automatic temperature recorder:  An automatic temperature recorder (optional accessory) can be mounted here.  See page 13.  4. Access port:    This port allows temperature measurement cables to enter the cham
                    
                    ページ11に含まれる内容の要旨 
                    
                           REFRIGERATOR COMPONENTS      Control panel and keypad    7 6 5                             1 2 3 4    1. Buzzer stop key (BZ):  To silence the audible alarm, press this key.  Press it once again to  reactivate the alarm.    2. Set key (SET):  Temperature setting mode is initiated by pressing this key.  Once the key is pressed,  the digit to be changed will flash.  Pressing this key again after setting the desired temperature stores  the  set temperature in the memory.    3. Digit shift key ( 
                    
                    ページ12に含まれる内容の要旨 
                    
                           INSTALLATION      Installation site    To operate this unit properly and to obtain maximum performance, install the unit in a location with the  following conditions:    1.  A location not subjected to direct sunlight  Installation in a location subjected to direct sunlight may lead to inadequate cooling.    2.  A location with adequate ventilation  Leave at least 10 cm around the unit for ventilation.  Poor ventilation will result in a reduction of the  refrigeration capacity.    3.  A locat
                    
                    ページ13に含まれる内容の要旨 
                    
                           INSTALLATION      Installation    1. Remove the packaging materials and tapes  Remove all transportation packaging materials and  tapes.  Open the doors and ventilate the unit.  If the  outside panels are dirty, clean them with a neutral  detergent and wipe it off with a soft cloth washed in    clean water.  Leveling foot  Fig. 1    2. Adjust the leveling foot  Extend the leveling feet by rotating them  counterclockwise until they make contact with the floor.  See Fig. 1.  Ensure the unit is 
                    
                    ページ14に含まれる内容の要旨 
                    
                           OPTIONAL COMPONENTS      Automatic temperature recorder    An automatic temperature recorder is available for this refrigerator as an optional accessory. If an  automatic temperature recorder is required, contact your Sanyo dealer.  For the proper usage of temperature recorder, refer to an instruction manual included with the recorder.  The available recorder and a dedicated fixture is as follows:  Recorder:  Model MTR-G04A (AC 100 to 150 V)  Model MTR-G04C (AC 220 to 240 V)  Fixture:  Model 
                    
                    ページ15に含まれる内容の要旨 
                    
                           OPTIONAL COMPONENTS      Attachment of recorder MTR-G04       WARNING Always disconnect the power supply plug before installing an automatic temperature recorder in order  to prevent electric shock or injury.    1. Follow the instruction manual provided with the  temperature recorder and attach the recorder to the  fixture MPR-S7.  shaft  screw    2. Open the cover for recorder attachment space by  cover  removing the screw.      Fig. 1  3. Remove the cover by pushing the shaft on both side  
                    
                    ページ16に含まれる内容の要旨 
                    
                           START-UP OF UNIT    Follow the procedures for the initial and consequent operations of the unit.   1. Connect the power cord to the dedicated outlet with appropriate rating.    2. On start-up, the alarm buzzer sometimes operates.  In this case, stop the buzzer by pressing the  alarm stop key (BZ).    o 3. Set the chamber temperature to 5 C.    o 4. Allow the chamber temperature to fall to 5 C.  Check the chamber temperature on the temperature  indicator.    5. Turn on the fluorescent light sw
                    
                    ページ17に含まれる内容の要旨 
                    
                           TEMPERATURE SETTING      Chamber temperature    Table 1 shows the basic procedure for setting the chamber temperature.  Perform key operations in the  sequence indicated in the table.  The example in the table is based on the assumption that the desired  o temperature is 4 C.  o Note:  The unit is set at the factory with a chamber temperature of 5 C.    o Table 1.  Basic operation sequence (Example:  Chamber set temperature 4 C)   Description of operation Key operated Indication after operati
                    
                    ページ18に含まれる内容の要旨 
                    
                           TEMPERATURE SETTING      Alarm temperature setting    This unit is provided with both high and low temperature alarms. The temperature at which the alarm is  activated may be changed.  o o o o The available set range for high temperature alarm is between +2 C and +14 C and -2 C and -14 C for  low temperature alarm against the chamber temperature.  o Note:  The temperature alarm is set at  5 C of the set temperature at the factory.  Display Mode Function   High temperature alarm set See Table 
                    
                    ページ19に含まれる内容の要旨 
                    
                           ALARMS & SAFETY FUNCTIONS    This unit has the alarms and safety functions shown in Table 5, and also self diagnostic functions.    Table 5.  Alarms and safety functions  Alarm & Safety Situation Indication Buzzer Safety operation  If the chamber temperature deviates ALARM lamp is flashed.  High temperature Intermittent tone with Remote alarm with 15  o from the set temperature +2 C or up to Temperature indicator is  alarm 15 minutes delay. minutes delay.  o +14 C. flashed.  If the chamber te
                    
                    ページ20に含まれる内容の要旨 
                    
                           ALARMS & SAFETY FUNCTIONS      Remote alarm terminal      WARNING  Always disconnect the power supply to the unit prior to connection of alarm equipment to the  terminal in order to prevent electric shock or injury.    The terminal to which a lead wire of remote alarm is connected is positioned in the box at the rear of the  cabinet.  The contact capacity is DC30V, 2A.  Contact output:  at normal condition;  “open”  at abnormal condition;  “close”    Note:  The alarm is silenced by the alarm