ページ1に含まれる内容の要旨 
                    
                         
OWNER’S MANUAL 
 
 
CO Meter 
2
Model IAQ10 
 
 
 
 
                                                                                                                                                                                                                                                                                                                                                                                                                                                             
                    
                    ページ2に含まれる内容の要旨 
                    
                        LIMITATION OF WARRANTY AND LIABILITY  Seller warrants the goods sold hereunder, under normal use and service as described in  the operator's manual, shall be free from defects in workmanship and material for  twenty-four (24) months, or the length of time specified in the operator's manual, from  the date of shipment to the customer. This warranty period is inclusive of any statutory  warranty.  This limited warranty is subject to the following exclusions:  a. Hot-wire or hot-film sensors used w
                    
                    ページ3に含まれる内容の要旨 
                    
                        Table of Contents  LIMITATION OF WARRANTY AND LIABILITY ...... ii  SECTION 1 General Description .....................................2  SECTION 2 Theory of Operation: NDIR Sensor ...........2  SECTION 3 Safety ..............................................................3  SECTION 4 Setting-Up the Model IAQ10 ......................4  Supplying Power ...........................................................4  Installing the Batteries...................................................4  Using 
                    
                    ページ4に含まれる内容の要旨 
                    
                        SECTION 1  General Description  The Model IAQ10 is a hand-held meter to  measure Carbon Dioxide (CO ), displayed  2 in units of parts per million (ppm). An  NDIR sensor in the top end of the meter  measures gas content by diffusion through  sensing holes.     The IAQ10 ships in a pouch that provides a  small amount of protection to the display  and the sensing holes in the top of the  instrument case. The instrument ships with  batteries (which are not installed), a  calibration sheet, and this 
                    
                    ページ5に含まれる内容の要旨 
                    
                        concentrations. This technique is called  non-dispersive infrared (NDIR) detection.      SECTION 3  Safety  When using the IAQ10 to check for CO  2 values, make certain that you can safely  raise and hold the instrument while  making measurements. Be especially  careful when working on a ladder.    Observe all necessary precautions so that  the unit does not become caught in moving  machinery or touch any exposed electrical  wiring.    DANGER!  Use with corrosive or other dangerous or  explosive
                    
                    ページ6に含まれる内容の要旨 
                    
                        SECTION 4  Setting-Up the Model IAQ10    Supplying Power  The IAQ10 can be powered in one of two  ways: four AA-size batteries or the optional  AC adapter.    Installing the Batteries  Insert four AA batteries as indicated by the  diagram located on the inside of the battery  compartment. The Model IAQ10 is de- signed to operate only with alkaline  batteries. When 15% battery life is remain- ing, the battery indicator will blink,  indicating the batteries need to be changed.  At 0%, “LO” will di
                    
                    ページ7に含まれる内容の要旨 
                    
                        SECTION 5  Operation of the Model IAQ10    Note: Do not hold the instrument close to your  face. Humans exhale CO and this will affect  2 the accuracy of the readings.     When pressing the keys on the keypad, the  Model IAQ10 will beep to confirm the  function. If you press a key and the Model  IAQ10 does not beep, the Model IAQ10  does not allow that function during the  selected mode. (See Section 9 for informa- tion on disabling this feature through an  internal DIP switch.)    ON/OFF Key  P
                    
                    ページ8に含まれる内容の要旨 
                    
                        calibration. When the Model IAQ10  completes the internal self-check, it will  display the approximate percentage of  battery life remaining.     Note: The sensor needs up to 10 minutes to  stabilize in still air before displayed readings  can be considered accurate. Moving the  instrument may decrease this stabilization time.    READ Key  Carbon dioxide concentrations will be  displayed in parts per million (ppm). The  IAQ10 can be left in a room to monitor  general conditions, but it is import
                    
                    ページ9に含まれる内容の要旨 
                    
                        The display updates every second and each  reading is the average of five samples.    ( 5) Key  The instrument only displays CO .   2   RECALL ( 6) Key  Press the RECALL key repeatedly to  display: minimum value (MIN), maximum  value (MAX), average value (AVG), and  number of seconds sampled (COUNT).  Press the READ key to return to normal  measuring mode.     The instrument does not store readings.  The information viewed in the recall mode  is for the sample reading just taken. Once a  new sam
                    
                    ページ10に含まれる内容の要旨 
                    
                        SECTION 6  Maintenance of the Model IAQ10  The Model IAQ10 requires very little  maintenance to keep it performing well.    Case  If the instrument case or storage case needs  cleaning, wipe it off with a soft, damp  cloth. Never submerge the IAQ10 in any  liquid or allow any liquid to enter the  sensing holes.    Storage  When storing the Model IAQ10 for more  than a month, TSI recommends removing  the batteries. This prevents possible  damage due to battery leakage.      SECTION 7  Service Inf
                    
                    ページ11に含まれる内容の要旨 
                    
                        Before sending your instrument for  calibration or repair, you should call  Customer Service. The service department  will provide you with the cost of service or  calibration, Return Material Authorization  (RMA) number, and shipping instructions.  Please have the following information  available when you call:  • Owner’s name, address, and phone  number  • Billing address, if different and  applicable  • Instrument name and model  • Serial number  • Date of purchase  • Where purchased  TSI rec
                    
                    ページ12に含まれる内容の要旨 
                    
                        you with a NIST (US National Institute of  Standards and Technology) traceable  certificate. This “annual checkup” assures  you of consistently accurate readings; it is  especially important in applications where  strict calibration records must be main- tained.    Send the instrument to TSI Instruments  Ltd. prepaid. Securely package your  instrument in a strong container sur- rounded by at least 5 cm of suitable shock- absorbing material. Include a purchase  order that clearly shows the instru
                    
                    ページ13に含まれる内容の要旨 
                    
                        1. Turn the instrument off. Under the  batteries, move DIP switch # 7 to ON.  Turn the instrument back on.  2. Press and hold the READ key to begin  the calibration procedure. The display  will begin a countdown from 5 to 0.  Release the key when the display reads  0. If the key is released too soon or too  late, the instrument will go back to sam- pling in real-time. “ZerO” will appear  on the display if the instrument is in  calibration mode.  3. Install the regulator on the zero calibra- tion
                    
                    ページ14に含まれる内容の要旨 
                    
                        indicates “SPAN” and the span concen- tration.  6. Install the regulator on the span calibra- tion cylinder and connect tubing from  the regulator to the inlet fitting on the  top of the Model IAQ10.  7. Use 5 and 6 to adjust the concentra- tion displayed on the Model IAQ10 to  match the concentration on the span gas  cylinder.  8. Press and release the READ key to take  a span gas reading. The Model IAQ10  starts a 60-second countdown. When the  countdown reaches zero, the instrument  returns t
                    
                    ページ15に含まれる内容の要旨 
                    
                        SECTION 8  Troubleshooting of the Model IAQ10  This table lists the symptoms, possible  causes, and recommended solutions for the  Model IAQ10.     Possible Corrective  Symptom Causes Action(s)  No display Unit not Switch on the  switched on unit   Low or dead Replace the  batteries batteries   Dirty battery Clean the  contacts battery  contacts   Batteries Refer to  installed battery  incorrectly illustration  inside battery  cover  BAT is Dirty battery Clean the  blinking contacts  battery  Ba
                    
                    ページ16に含まれる内容の要旨 
                    
                         Possible Corrective  Symptom Causes Action(s)  Display Batteries are Replace the  reads “LO” low batteries    Note: If your symptoms are not remedied  by the suggested corrective action, call TSI  Instruments Ltd.       SECTION 9  Internal DIP Switch Settings  To access the DIP switches, remove the  batteries from the battery compartment. On  the inside of the battery compartment,  there is a window with eight DIP switches.  The table below shows the functions for  each switch.     CAUTION:  Ma
                    
                    ページ17に含まれる内容の要旨 
                    
                        DIP    Switch OFF ON  1-3 Must be OFF -----  4 not assigned  not assigned  5 not assigned not assigned  6 Beep is turned Beep is turned  OFF ON   7 Normal mode Field Verifica- tion mode  8 not assigned not assigned  - The ON position is away from the batteries  and OFF is towards the batteries.  - Switch 1 is towards the top of the case and  Switch 8 is nearest the bottom.  15                                                                                                                         
                    
                    ページ18に含まれる内容の要旨 
                    
                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            
                    
                    ページ19に含まれる内容の要旨 
                    
                        Specifications   (subject to change without notice)    CO Sensor type:  2  Dual wavelength detector with non- dispersive infrared (NDIR) sensor    Range: 0 to 5000 ppm    Accuracy ±3% of reading or ±50 ppm, whichever  1, 2 is greater    Resolution 1 ppm    Response time <10 minutes in still air    Operating Temperature Range:   5 to 70°C    External Meter Dimensions:    89 mm × 168 mm × 41 mm    Display: LCD, 4 digits, 10-mm high    Power requirements:   Four (4) AA-size alkaline batteries    Ba
                    
                    ページ20に含まれる内容の要旨 
                    
                                                                 TSI Instruments Ltd.  October 2005  Lancaster Road Printed in USA  Cressex Business Park Part No. 1980532, Rev. A  High Wycombe   Buckinghamshire Tel:  +44 (0) 149 4 459200  HP12 3QP  England Fax: +44 (0) 149 4 459700     E-mail: info@airflowinstruments.co.uk  Copyright © 2005  Web: www.airflowinstruments.co.uk  TSI Incorporated