ページ1に含まれる内容の要旨 
                    
                         
 
 
 
 
IDEAL INDUSTRIES INC. 
TECHNICAL MANUAL 
MODELS: 61-633 
61-635 
 
 
 
The Service Information provides the following information: 
•   Precautions and safety information 
•   Specifications     
•   Basic maintenance (cleaning, replacing the battery and fuses) 
•   Performance test procedures 
•   Calibration and calibration adjustment procedures 
 
 
 
 
Form Number:  TM61633-5 
Revision: 5. Date:  December 2006 
 
Form Number TM61633-5 Rev 5 Dec 2006                                 
                    
                    ページ2に含まれる内容の要旨 
                    
                        TABLE OF CONTENTS  Title                                                                                          Page 1 Introduction  1 Precautions and Safety Information  Symbols 1 Safety Information 2 3 Specifications  3      General Specification        Voltage Specifications  4 4      dBm / dB Specifications  5      Current Specifications  5       Peak Hold, Resistance Specifications        Continuity and Diode Specifications 5      Capacitance Specifications 6 6      Frequency Counter and 
                    
                    ページ3に含まれる内容の要旨 
                    
                        Page 1    Introduction  Warning  To avoid shock or injury, do not perform the verification tests or calibration  procedures described in this manual unless you are qualified to do so.  The information provided in this document is for the use of qualified personnel only.     Caution  The 61-630 serials contain parts that can be damaged by static discharge.  Follow the standard practices for handling static sensitive devices.    For additional information about IDEAL INDUSTRIES and its products,  
                    
                    ページ4に含まれる内容の要旨 
                    
                        Page 2    SAFETY    Review the following safety precautions to avoid injury and prevent damage to this product or any products  connected to it. To avoid potential hazards, use the product only as specified.     CAUTION.   These statements identify conditions or practices that could result in damage to the equipment or other property.     WARNING.   These statements identify conditions or practices that could result in personal injury or loss of life.    Specific precautions    Use proper Fuse. 
                    
                    ページ5に含まれる内容の要旨 
                    
                        Page 3    SPECIFICATIONS  All specifications are warranted unless noted typical and apply to models 61-633 and 61-635.    °  Stated accuracies are at 23 ±5°C at less than 80% relative humidity and without the battery indicator displayed.    General specifications    Characteristics Description  LCD display digits 4¾ or 3¾  Bargraph segments 80 Segment Graph  Display count 40,000 or 4,000  Numeric update rate 2 times / sec (40.000 count)  4 times / sec (4,000 count)  Bargraph 20 times/sec  Polari
                    
                    ページ6に含まれる内容の要旨 
                    
                        Page 4    Measurement Characteristics   All at 23°C ±5°C, < 80% Rh.   Specifications are expressed as ±([% of reading] + [number of digits]).  Multiply accuracy Digits by l0 in 40,000 count Mode.    DC VOLTAGE:  DCV 61-633 61-635  40mV ± (0.20%+8d) ± (0.06% + 8d)  400mV ± (0.20% + 2d) ± (0.06% + 2d)  4V, 40V, 400V, 1000V ± (0.20% + 2d) ± (0.06% + 1d)    AC VOLTAGE:  ACV 61-633 61-635  400mV    40Hz~ 100Hz ± (1.20% + 5d) ± (0.70% + 5d)  100Hz ~ 1KHz ± (2.00% + 5d) ± (1.00% + 5d)  4V    40Hz ~ 100
                    
                    ページ7に含まれる内容の要旨 
                    
                        Page 5    DC CURRENT:  DCA 61-633 61-635  40mA, 400mA, 4A, 10A ± (0.50% + 4d) ±(0.20% + 4d)    AC CURRENT:  ACA 61-633 61-635  40mA, 400mA, 4A, 10A ±(1.20% + 8d) ± (0.80% + 8d)  Bandwidth 40Hz ~ 400Hz 40Hz ~ 400Hz    Range:                 40mA, 400mA, 4A, 10A  Resolution:          1µA in the 40mA range  Burden Voltage:  800mV max. for mA input, 1V max. for A input  AC Conversion Type: AC Coupled True RMS responding  Input Protection:  Equipped with High Energy Fuse                         1A, 6
                    
                    ページ8に含まれる内容の要旨 
                    
                        Page 6    CAPACITANCE  Capacitance 61-633 61-635  4nF, 40nF  ± (1 .90% + 20d) ± (0.90% + 20d)  400nF, 4µF  40µF, 400µF ± (2.90% + 20d) ± (1 .90% + 20d)  4mF, 10mF ± (3.90% + 20d) ± (2.90% + 20d)    Note:   For best measurements, use REL  mode in the nF ranges.  Range: 4nF, 40nF, 400nF, 4µF, 40µF, 400µF, 4mF, l0mF  Resolution: lpF in the 4nF range  Input Protection: 600V rms    FREQUENCY COUNTER  Range:  400Hz, 4KHz, 40KHz, 400KHz, 4MHz  Resolution:  0.01Hz in the 400Hz range  Accuracy:  ± (0.01%
                    
                    ページ9に含まれる内容の要旨 
                    
                        Page 7    Certifications and compliances  Safety   Designed to IEC 1010-1, UL3111-1 and CSA specifications    1000V DC Category II    600V DC Category III  Input rating    750V AC Category II    600V AC Category III    CAT III: Distribution level mains, fixed installation.    CAT II: Local level mains, appliances, portable equipment.  Over voltage category    CAT I: Signal level, special equipment or parts                    of equipment, telecommunication, electronics.    Do not operate in envi
                    
                    ページ10に含まれる内容の要旨 
                    
                        Page 8    Required Equipment            Required equipment is listed in Table B. If the recommended             models are not available, equipment with equivalent specifications may be used.            Repairs or servicing should be performed only by qualified personnel.    Table B. Required Equipment  Recommended  Equipment Required Characteristics  Model  Calibrator AC Voltage Range: 0-750V ac Fluke 5500 or  Accuracy: ±0.07% (Basic) Wavetek 9100  Frequency Range: 40 ~ 100KHz Calibrator or   A
                    
                    ページ11に含まれる内容の要旨 
                    
                        Page 9    Basic Maintenance   Warning            To avoid shock, remove the test leads and any input signals before opening             the case or replacing the battery or fuses.    Opening the Meter Case  Caution            To avoid unintentional short circuit, always place the uncovered Meter             assembly on a protective surface. When the case of the Meter is open, circuit             connections are exposed.    To open the Meter case, refer to Figure 1 and do the following:  1.   Dis
                    
                    ページ12に含まれる内容の要旨 
                    
                          Page 10          Form Number TM61633-5 Rev 5 Dec 2006                                                                                                                                                                                                                                                                                                                                                                                                                                                             
                    
                    ページ13に含まれる内容の要旨 
                    
                        Page 11    Replacing Fuses   Warning         To avoid electrical shock, remove the test leads and any input signals before          replacing the battery or fuses. To prevent damage or injury, INSTALL ONLY          quick acting fuses with the following Amp/Volt current interrupt rating:         FS1 Fuse: 1A, 600V, FAST. Minimum interrupt rating 10,000A         FS2 Fuse: 15A, 600V, FAST. Minimum interrupt rating 100,000A     To replace the Meter’s fuses, refer to Figure 1 and do the following:  1
                    
                    ページ14に含まれる内容の要旨 
                    
                        Page 12    Performance Tests  The following performance tests verify the complete operability of the   Meter and check the accuracy of each Meter function against the Meter’s specifications.  Accuracy specifications are valid for a period of one year after calibration, when measured at an operating  temperature of 18°C to 28°C and at a maximum of 80% relative humidity.    To perform the following tests, it is not necessary to open the case; no adjustments are necessary. Merely  make the required
                    
                    ページ15に含まれる内容の要旨 
                    
                        Page 13    Testing the Voltage Function  To verify accuracy in the DC and AC voltage ranges, do the following:  1.   Connect the Calibrator to the V Ω and COM inputs on the Meter.  2.   Set the Calibrator for the voltage from step 1 to 5 in Table 1.        Compare the reading on the Meter display with the display reading shown in Table 1.        If the display reading falls outside of the range shown in Table 1, the Meter does not         meet specification.    Table 1. DC Voltage Test  Input Re
                    
                    ページ16に含まれる内容の要旨 
                    
                        Page 14    Testing Millivoltage (mV) Function  To verify accuracy of the mV function, do the following.  1.   Connect the Calibrator to the V Ω  and COM inputs on the meter.  2.   Turn the rotary switch to mV.  3.   Set the calibrator for the voltage from step 1 to 3 in Table 3.  4.   Compare the reading on the Meter display with the display reading in Table 3.        If the display reading falls outside of the range shown in Table 3, the Meter does not        meet specification.    Table 3. DC 
                    
                    ページ17に含まれる内容の要旨 
                    
                        Page 15    Testing the Resistance Function  To verify the accuracy of the resistance function, do the following:  1.   Connect the Calibrator to V Ω and COM on the Meter.  2.   Turn the rotary switch to Ω  3.   Apply the inputs for step 1-7 in Table 5.        *Set reference if using a Multifunction Calibrator with diff or zero mode for steps 1, 2, and 3.        Compare the Meter display readings to the display readings in Table 5 step 1-7.  4.   If the display reading falls outside of the range 
                    
                    ページ18に含まれる内容の要旨 
                    
                        Page 16    Testing the Capacitance Function (cont’d)        d)  For each input, compare the readings on the Meter display to display readings in               Table 7. The meter selects the proper range automatically. Each measurement takes               about one second per range. 5mF will taken about 15 seconds.  3.   If the display reading falls outside of the range shown in Table 7, the Meter does not        meet specification.    Table 7. Capacitance Test  Input Readings  Step Capacitance 6
                    
                    ページ19に含まれる内容の要旨 
                    
                        Page 17    Table 9. AC mA Test  Input Readings  Step Current Frequency 61-633 61-635  1 36.000mA 50 Hz 35.488 to 36.512 35.632 to 36.368  2 36.000mA 400 Hz 35.488 to 36.512 35.632 to 36.368  3 360.00mA 50 Hz 354.88 to 365.12 356.32 to 363.68  4 360.00mA 400Hzz 354.88 to 365.12 356.32 to 363.68      Testing the Amp (A) Function  To verify the accuracy in the ampere (A) measurement function, do the  Following:  1.   Connect the Calibrator to the A and COM inputs of the Meter.  2.   Turn the rotary
                    
                    ページ20に含まれる内容の要旨 
                    
                        Page 18    Table 12. Frequency Test  Input Reading  Step Frequency Level 61-633 61-635  1 360Hz 0.5Vp-p 359.86 to 360.14 359.86 to 360.14  2 3.6000KHz 0.5Vp-p 3.5986 to 3.6014 3.5986 to 3.6014  3 36.000KHz 0.5Vp-p 35.986 to 36.014 35.986 to 36.014  4 360.00KHz 0.5Vp-p 359.86 to 360.14 359.86 to 360.14  5 3.6000MHz 1.5Vp-p 3.5986 to 3.6014 3.5986 to 3.6014    Testing the Duty Factor Function  To verify the accuracy of the Meter’s Duty Factor function, do the following:  1.   Connect the Calibrato