Summary of the content on the page No. 1 
                    
                        TELEDYNE  
HASTINGS 
INSTRUMENTS 
INSTRUCTION MANUAL 
201/203/205/207 SERIES 
FLOW METERS/CONTROLLERS 
ISO 9001
CER T IFIED                                                                                                                                                                                                                                                                                                                                                                                         
                    
                    Summary of the content on the page No. 2 
                    
                        Manual Print History  The print history shown below lists the printing dates of all revisions and addenda created for this  manual. The revision level letter increases alphabetically as the manual undergoes subsequent  updates. Addenda, which are released between revisions, contain important change information that  the user should incorporate immediately into the manual. Addenda are numbered sequentially.  When a new revision is created, all addenda associated with the previous revision of the 
                    
                    Summary of the content on the page No. 3 
                    
                        Table of Contents    1. GENERAL INFORMATION ............................................................................................................................................ 4  1.1. FEATURES .................................................................................................................................................................... 4  1.2. SPECIFICATIONS HFM-201/205 ..................................................................................................
                    
                    Summary of the content on the page No. 4 
                    
                        1. General Information  The Hastings HFM-201/HFM-205/HFC-203/HFC-207 series Mass flow meters (HFM-201/HFM- 205) and controllers (HFC-203/HFC-207) are designed to accurately measure and control mass  flow over the range of 30 slm to 2500 slm, without corrections or compensations for gas pressure  and temperature with an accuracy of better than ±1% from the mean (±2% FS for 201/203 units in  the ½” fitting size base with flows >500 slm).  Hastings mass flow instruments do not require any  periodic
                    
                    Summary of the content on the page No. 5 
                    
                        1.2. Specifications HFM-201/205   1 Accuracy and Linearity ....................................................................................................... ±1% F.S.  Repeatability  .................................................................................................................. ±0.05% F.S.   Standard Pressure Rating ...................................................................................................... 500 psig  Pressure Coefficient  .......................
                    
                    Summary of the content on the page No. 6 
                    
                        1.4. Optional 4 - 20 mA Current Output  An option to the standard 0 - 5 VDC output is the 4 - 20 mA current output that is proportional to  flow.  The 4 - 20 mA signal is produced from the 0 - 5 VDC output of the flow meter.  The current  loop output is useful for remote applications where pickup noise could substantially affect the  stability of the voltage output or long cable runs where cable resistance would cause a voltage signal  to decay.  The current loop signal replaces the voltage outp
                    
                    Summary of the content on the page No. 7 
                    
                        feature.  More information about Hastings Instruments’ power supplies can be found on the  Hastings web site.  http://www.teledyne-hi.com/products/powerpod-series.htm  1.5.2. Interconnecting Cables  Cables are available from Hastings, in various lengths, to connect from the 15 pin "D" connector on  the back of Hastings power supplies directly to any of the 200 series and 300 series flow instruments  (including digital versions).   More information about the available cables can be found in the T
                    
                    Summary of the content on the page No. 8 
                    
                        2. Installation and Operation  This section contains the necessary steps to get a new flow meter/controller into operation as quickly  and easily as possible.  Please read the following thoroughly before attempting to install the  instrument.  2.1. Receiving Inspection  Carefully unpack the Hastings 201/203 series instrument and any accessories that have also been  ordered.  Inspect for any obvious signs of damage to the shipment.  Immediately advise the carrier  who delivered the shipment if an
                    
                    Summary of the content on the page No. 9 
                    
                        2.4. Mechanical Connections  The flow meter may be mounted in any position as long as the direction of gas flow through the  instrument follows the arrow marked on the bottom of the flow meter case label.  The preferred  orientation is with the inlet and outlet fittings in a horizontal plane (if operating with a dense gas or at  high pressures the instrument must be installed horizontally).  When mounted in a different  orientation the instrument should be re-zeroed at zero flow with the system 
                    
                    Summary of the content on the page No. 10 
                    
                        the ground connection at the power supply. Do not connect them together at the flow controller as  the resulting crosstalk could result in flow instabilities.   24 Volt Connections  Refer to the diagram at right when connecting 24 Volt  units.  Connect the positive lead of the power supply to pin 7 of  the DE-9 connector and negative lead to pin 4.  The  supply input is diode protected such that reversing the  input polarity will not damage the instrument.   The  power supply is galvanically iso
                    
                    Summary of the content on the page No. 11 
                    
                        some zero offset.)  The 201/203 series is intended for use in non-condensing environments only.   Condensate or any other liquids which enter the flow meter may destroy its electronic components.  2.6.2. Zero Check  Turn the power supply on if not already energized.  Allow for a 1 hour warm-up.  Stop all flow  through the instrument and wait 2 minutes.  Caution:  Do not assume that all metering valves  completely shut off the flow.  Even a slight leakage will cause an indication on the meter and
                    
                    Summary of the content on the page No. 12 
                    
                        2.6.5. Operation with a Hastings power supply.  There are two controls for each flow controller connected to a Hastings power supply.  A switch  labeled “OPEN; AUTO; CLOSED” (valve override THPS 400 only) and a potentiometer labeled  “COMMAND”.  For normal operation, the valve override switch will be in the “AUTO” position.   The “CLOSE” position removes all power from the valve, shutting off flow regardless of the  command pot setting.  The “OPEN” position applies full available valve voltage t
                    
                    Summary of the content on the page No. 13 
                    
                        2.6.7. Operation with an external sensor.  (Fig.  2.2)  In some instances, it might be desirable to use an external sensor to provide process information to  the control circuitry in the flow controller.  For example, you might want to control the pressure in a  vacuum system by adjusting the rate at which the system is backfilled with a gas.  The new,  enhanced HFC series of flow controllers have provision for accepting a 0 - 5VDC (or 0 - 10 VDC)  output from an external sensor at pin 13 of the
                    
                    Summary of the content on the page No. 14 
                    
                        3) Step change the command voltage to 100%, and observe the flow through the controller.  If the overshoot  is too large, remove the jumper.  Reset the command voltage to 10%, and allow the controller to stabilize.  4) To prevent loss of the unused jumper, place it over one pin only on JP4/JP6.  2.6.9. Operating Temperature  For proper operation, the combination of ambient temperature and gas temperature must be such  that the Flowcontroller temperature remains between 0 and 50°C.  Most accurate
                    
                    Summary of the content on the page No. 15 
                    
                        3. Theory of Operation  This section contains an overall functional description of HFC Flow Controllers.  Detailed  schematics and parts lists can be found at the end of the manual in Section 6.0.  In this section and  other sections throughout this manual, when a power supply is mentioned, it is assumed that the  customer has a Hastings Power Supply.  These sections are not applicable if another type of power  supply is used.  3.1. Overall Functional Description:   The HFC Flow Controller consi
                    
                    Summary of the content on the page No. 16 
                    
                        output is sent to the terminals on the back  and to the decoding circuitry in the display  which converts it to a 3-digit output.  The controller circuitry utilizes the  Command and the Flow voltages as input  signals.  The 0 - 5VDC command signal is  subtracted from the 0 - 5VDC flow signal  creating an error signal.  This signal is  amplified and causes the solenoid valve to  move.  The amount and direction of the  movement is dependent upon the value and  the sign of the error signal, and ten
                    
                    Summary of the content on the page No. 17 
                    
                        3.5. Valve:  A solenoid valve is used as a pilot valve to control a much larger pneumatic diaphragm valve.   The  pilot valve is an “automatic metering solenoid” valve.  While most solenoids operate in either the  fully open or fully closed state, the automatic metering solenoid valve is designed to control flow (see  Figure 3.5).  A spring, connected to the plunger assembly, holds a magnetic plunger tightly against  an orifice to shut off flow.  The magnetic plunger is surrounded by an electric
                    
                    Summary of the content on the page No. 18 
                    
                        4. Maintenance  This section contains service and calibration information.  Some portions of the instrument are  delicate.  Use extreme care when servicing the flow controller. The potentiometer positions and the  electrical components referred to in the troubleshooting section can be found in Section 6.0 on the  electrical component layout drawing.  4.1. Authorized Maintenance  With proper care in installation and use, the flow controller will require little or no maintenance.  If  maintenance 
                    
                    Summary of the content on the page No. 19 
                    
                        restriction is pneumatically close to the downstream end of the flow controller.  The differential  pressure across the unit must be between 10-50 psig.    SYMPTOM:  Little or no flow, even with Valve Override switch OPEN.  CAUSE:  Plugged orifice.  ACTION:  Verify the presence of a 10-50 psig pressure across the instrument.  If present, shut off  gas supply and power supply.  Remove orifice per Section 4.9.  Examine orifice.  If plugged, clean or  replace as applicable.  Reassemble valve.    SY
                    
                    Summary of the content on the page No. 20 
                    
                        stability.  If the instrument is not shutting completely off when Valve Override switch is in the  CLOSE position, the orifice may require approximately 1/8 turn clockwise.  4.4. End Cap Removal:  The end cap on the inlet side must be removed to gain access to the filter or shunt assembly. First  shut off the supply of gas to the instrument.  Disconnect the Swagelok fitting on the inlet and outlet  sides of the transducer, and remove it from the system plumbing.  Remove the four hex bolts holdin