ページ1に含まれる内容の要旨 
                    
                        OPERATING INSTRUCTIONS AND SPECIFICATIONS
NI 9201/9221
8-Channel, 12-Bit Analog Input Modules
Français Deutsch
ni.com/manuals                                                                                                                                                                                                                                                                                                                                                                                       
                    
                    ページ2に含まれる内容の要旨 
                    
                        This document describes how to use the National Instruments 9201  and National Instruments 9221 and includes specifications and  terminal assignments for the NI 9201/9221. In this document,  the NI 9201/9221 with screw terminal and NI 9201/9221 with  DSUB are referred to inclusively as the NI 9201/9221. Visit  ni.com/info and enter rdsoftwareversion to determine  which software you need for the modules you are using. For  information about installing, configuring, and programming the  system, re
                    
                    ページ3に含まれる内容の要旨 
                    
                        Safety Guidelines Operate the NI 9201/9221 only as described in these operating  instructions. Hot Surface This icon denotes that the component may be  hot. Touching this component may result in bodily injury. Safety Guidelines for Hazardous Voltages You can connect hazardous voltages only to the NI 9201/9221 with  screw terminal. Do not connect hazardous voltages to the  NI 9201/9221 with DSUB. If hazardous voltages are connected to the module, take the  following precautions. A hazardous volta
                    
                    ページ4に含まれる内容の要旨 
                    
                        Caution Make sure that devices and circuits connected to  the module are properly insulated from human contact. Caution When module terminals are hazardous voltage  LIVE (>42.4 V /60 VDC), you must ensure that devices  pk and circuits connected to the module are properly  insulated from human contact. You must use the NI 9932  connector backshell kit to ensure that the terminals are  not accessible. Figure 1 shows the NI 9932 connector backshell. Note You can use the NI 9932 connector backshell 
                    
                    ページ5に含まれる内容の要旨 
                    
                        Figure 1.  NI 9932 Connector Backshell Safety Guidelines for Hazardous Locations The NI 9201/9221 is suitable for use in Class I, Division 2,  Groups A, B, C, D, T4 hazardous locations; Class I, Zone 2,  AEx nC IIC T4, and Ex nC IIC T4 hazardous locations; and  nonhazardous locations only. Follow these guidelines if you are  installing the NI 9201/9221 in a potentially explosive environment.  Not following these guidelines may result in serious injury or  death. © National Instruments Corp. 5 NI
                    
                    ページ6に含まれる内容の要旨 
                    
                        Caution Do not disconnect I/O-side wires or connectors  unless power has been switched off or the area is known  to be nonhazardous. Caution Do not remove modules unless power has been  switched off or the area is known to be nonhazardous. Caution Substitution of components may impair  suitability for Class I, Division 2. Caution For Zone 2 applications, install the system  in an enclosure rated to at least IP 54 as defined by  IEC 60529 and EN 60529.  Caution For Zone 2 applications, install a 
                    
                    ページ7に含まれる内容の要旨 
                    
                        locations. If you are using the NI 9201/9221 in Gas Group IIC  hazardous locations or in ambient temperatures of  –40 °C ≤ Ta ≤ 70 °C, you must use the device in an NI chassis  that has been evaluated as EEx nC IIC T4, Ex nA IIC T4, or  Ex nL IIC T4 equipment. Special Conditions for Marine Applications Some modules are Lloyd’s Register (LR) Type Approved for  marine applications. To verify Lloyd’s Register certification, visit  ni.com/certification and search for the LR certificate, or  look for
                    
                    ページ8に含まれる内容の要旨 
                    
                        Connecting the NI 9201/9221 The NI 9201/9221 provides connections for eight analog input  channels. AI0 0 AI1 1 1 COM AI0 14 AI2 2 2 NC NC 15 3 COM AI3 3 AI1 16 4 COM AI2 17 AI4 4 5 NC NC 18 6 COM AI3 19 AI5 5 7 COM AI4 20 8 NC AI6 6 NC 21 9 COM AI5 22 AI7 7 10 COM AI6 23 11 NC NC 8 NC 24 12 COM AI7 25 13 COM COM 9 Figure 2.  NI 9201/9221 Terminal and Pin Assignments NI 9201/9221 8 ni.com                                                                                                             
                    
                    ページ9に含まれる内容の要旨 
                    
                        The NI 9201/9221 with screw terminal has a 10-terminal,  detachable screw-terminal connector. The NI 9201/9221 with  DSUB has a 25-pin DSUB connector. Each channel has an AI  terminal or pin to which you can connect a voltage signal. COM,  the common terminal or pin, is internally connected to the isolated  ground reference of the module. The NI 9201/9221 channels are isolated from other modules in the  system. The module protects each channel from overvoltages.  Refer to the Specifications sect
                    
                    ページ10に含まれる内容の要旨 
                    
                        You can connect single-ended voltage signals to the NI 9201/9221.  Connect the positive lead of the voltage signal to AI. Connect the  ground signal to COM. Refer to Figure 3 for an illustration of  connecting a single-ended voltage signal to the NI 9201/9221.   AI Voltage Source COM NI 9201/9221 Figure 4.  Connecting a Single-Ended Voltage Signal to the NI 9201/9221 Note You must use 2-wire ferrules to create a secure  connection when connecting more than one wire to a  single terminal on the N
                    
                    ページ11に含まれる内容の要旨 
                    
                        screw-terminal connector or use the NI 9932 backshell kit to  protect the connections. Refer to Figure 5 for an illustration of  using ferrules. Refer to Figure 1 for an illustration of the NI 9932  connector backshell. Figure 5.  10-Terminal Detachable Screw-Terminal Connector with Ferrule © National Instruments Corp. 11 NI 9201/9221                                                                                                                                                                    
                    
                    ページ12に含まれる内容の要旨 
                    
                        Sleep Mode This module supports a low-power sleep mode. Support for sleep  mode at the system level depends on the chassis that the module is  plugged into. Refer to the chassis manual for information about  support for sleep mode. If the chassis supports sleep mode, refer to  the software help for information about enabling sleep mode. Visit  ni.com/info and enter cseriesdoc for information about  C Series documentation. Typically, when a system is in sleep mode, you cannot  communicate with th
                    
                    ページ13に含まれる内容の要旨 
                    
                        Input Characteristics Number of channels..........................8 analog input channels ADC resolution.................................12 bits Type of ADC.....................................Successive approximation  register (SAR) Sample rate (aggregate) Maximum Sample Maximum  Rate (R Series Sample Rate  Module Expansion Chassis) (All Other Chassis) NI 9201, 475 kS/s 800 kS/s single channel NI 9201, scanning 475 kS/s 500 kS/s NI 9221 475 kS/s 800 kS/s Input range NI 9201.....................
                    
                    ページ14に含まれる内容の要旨 
                    
                        1 Operating voltage ranges Maximum Voltage,  Measurement Voltage, Channel-to-Earth Ground or  Channel-to-COM COM-to-Earth Ground Min Typ Max Screw Module (V) (V) (V) Terminal DSUB NI 9201 ±10.3 ±10.53 ±10.8 250 V ±60 VDC rms NI 9221 ±61.4 ±62.50 ±63.8 Overvoltage protection (channel-to-COM) ............................±100 V 1   Refer to the Safety Guidelines section for more information about safe operating  voltages. NI 9201/9221 14 ni.com                                                       
                    
                    ページ15に含まれる内容の要旨 
                    
                        NI 9201 accuracy (excludes noise) Percent of  Percent of  * Reading Range  Measurement Conditions (Gain Error) (Offset Error) Calibrated typ (25 °C, ±5 °C) ±0.04% ±0.07% Calibrated max (–40 to 70 ºC) ±0.25% ±0.25% Uncalibrated typ (25 °C, ±5 °C) ±0.26% ±0.46% Uncalibrated max (–40 to 70 ºC) ±0.67% ±1.25% *  Range equals 10.53 V © National Instruments Corp. 15 NI 9201/9221                                                                                                                              
                    
                    ページ16に含まれる内容の要旨 
                    
                        NI 9221 accuracy (excludes noise) Percent of  Percent of  * Reading Range  Measurement Conditions (Gain Error) (Offset Error) Calibrated typ (25 °C, ±5 °C) ±0.04% ±0.07% Calibrated max (–40 to 70 ºC) ±0.25% ±0.25% Uncalibrated typ (25 °C, ±5 °C) ±0.26% ±0.43% Uncalibrated max (–40 to 70 ºC) ±0.67% ±1.06% *  Range equals 62.50 V Stability Gain drift ....................................±34 ppm/°C Offset drift NI 9201 ................................±100 μV/°C NI 9221 ..............................
                    
                    ページ17に含まれる内容の要旨 
                    
                        Input impedance Resistance...................................1 MΩ  Capacitance ................................5 pF Input noise (code-centered) RMS ...........................................0.7 LSB rms Peak-to-peak...............................5 LSB No missing codes..............................12 bits DNL ..................................................–0.9 to 1.5 LSB INL....................................................±1.5 LSB Crosstalk (at 10 kHz)........................–75 dB Settling
                    
                    ページ18に含まれる内容の要旨 
                    
                        MTBF ...............................................1,485,465 hours at 25 °C;  Bellcore Issue 6, Method 1,  Case 3, Limited Part Stress  Method Note Contact NI for Bellcore MTBF specifications  at other temperatures or for MIL-HDBK-217F  specifications. Power Requirements Power consumption from chassis Active mode ...............................550 mW max Sleep mode .................................1 mW max Thermal dissipation (at 70 °C) Active mode ...............................600 mW max Slee
                    
                    ページ19に含まれる内容の要旨 
                    
                        Physical Characteristics If you need to clean the module, wipe it with a dry towel. Screw-terminal wiring ......................12 to 24 AWG copper  conductor wire with 10 mm  (0.39 in.) of insulation  stripped from the end 2 2 Ferrules .............................................0.25 mm to 2.5 mm Torque for screw terminals ...............0.5 to 0.6 N · m  (4.4 to 5.3 lb · in.) Weight NI 9201/9221  with screw terminal ....................150 g (5.3 oz) NI 9201/9221 with DSUB..........145 g (5.1
                    
                    ページ20に含まれる内容の要旨 
                    
                        NI 9201/9221 with Screw Terminal Isolation Voltages Channel-to-channel...........................No isolation between  channels Channel-to-earth ground ,  Continuous .................................250 V rms Measurement Category II Withstand....................................2,300 V , verified by a 5 s  rms dielectric withstand test  Measurement Category II is for measurements performed on  circuits directly connected to the electrical distribution system.  This category refers to local-level