LF654 /LF620 
Field Intelligent Device – Mount Anywhere Series - Flanged 
LF654 /LF622 
 Electromagnetic Flowmeter 
1/2" to 18" (15 to 450 mm) 
 
 
 
Introduction 
The electromagnetic flowmeter uses Faraday’s Law of 
electromagnetic induction to measure the process flow. 
The device consists of two units: a detector, through 
which the fluid to be measured flows and in which 
low-level signals proportional to flow rates are 
obtained; and a converter, which supplies excitation 
current to the 
                    
                    ページ2に含まれる内容の要旨 
                    
                        TIC-LF654L  Drinking water approvals:  Fluid conductivity:  PU and HR liner:  NSF certified   3µS/cm minimum     Fluid temperature:    Model LF654 Detector  -4 to 248 °F (-20 to +120 °C):   Mounting style: Flange connection type  PFA lined flowmeter   - Fluid pressure:     -4 to 140 °F ( 20 to +60 °C):   - - - 15 psi or 1.0 bar ( 0.1 MPa) to the nominal          Polyurethane lined flowmeter    - pressure of the connection flange.  23 to 176 °F ( 5to +80 °C):   Note: Before shipping from the man
                    
                    ページ3に含まれる内容の要旨 
                    
                        TIC-LF654L  Pulse width: 0.5 to 500ms (but less than half of    Model LF620 and LF622 converters  the period for 100% flow rate)  Input signals  Note: The same and simultaneous pulse is not available  Analog signal — the voltage signal from detector,  between DO1 and DO2.)  proportional to process flow rate (for LF622  • Multi-range selection outputs (Note 1)  separate type converter).  • High, High high, Low, and/or Low low alarm  Digital input DI   outputs (Note 2)  Signal type:  20 to 30Vdc 
                    
                    ページ4に含まれる内容の要旨 
                    
                        TIC-LF654L  LCD display:  Surge protection:   Arresters are installed in the power supply and a   Full dot-matrix 128×128 dot LCD display  current signal output circuit to help protect the meter  (back–light provided)  from lightning and improve personnel safety.  A parameter change will rotate the display.  Case: Aluminum alloy (equal to IP 67)  Parameter settings — Parameters can be set as  follows:  Coating:   • IR Switches: Three key switches are provided  Acrylic resin-baked coating, pearl–
                    
                    ページ5に含まれる内容の要旨 
                    
                        TIC-LF654L  Installation   Dimensions   Unit: inch (mm)    Note1: Eye bolts are provided at the top for flowmeters sized 8" (200mm) or above, and further, a roll-prevention base is provided for  flowmeters sized 10" (250mm) or larger.  Note2: Cable glands are not provided for LF654 of cFMus approved type.  Refer to the cable connection port at converter.    ASME B 16.5 class 150 dimensions: JIS B 2220 10K dimensions:  Meter size L1 L2 L3 No. of Weight Meter size L1 L2 L3 No. of Weight  (inch) (
                    
                    ページ6に含まれる内容の要旨 
                    
                        TIC-LF654L    Note1: Eye bolts are provided at the top for flowmeters sized 8" (200mm) or above, and further, a roll-prevention base is  provided for flowmeters sized 10" (250mm) or larger.    Note2: Cable glands are not provided for LF654 of cFMus approved type. Refer to the cable connection port at converter.    ASME B 16.5 class 150 dimensions: JIS B 2220 10K dimensions:  Unit: inch (mm)  Meter size L1 L2 L3 No. of Weight Meter size L1 L2 L3 No. of Weight  (inch) (inch) (inch) (inch) bolts (l
                    
                    ページ7に含まれる内容の要旨 
                    
                        TIC-LF654L                                                           Note1: Eye bolts are provided at the top for flowmeters sized 8”(200mm) or above, and further, a      roll-prevention base is provided for flowmeters 10”(250mm) or larger.  Note2: Cable glands are not provided for LF654 of cFMus approved type. Refer to the cable connection     d JIS B 2220 10K dimensions:  ASME B 16.5 class 150 dimensions:  Meter size L1 L2 L3 No. of Weight Meter size L1 (L2) L3 No. of Weight  (inch) (inch) (in
                    
                    ページ8に含まれる内容の要旨 
                    
                        TIC-LF654L                                            Note1: Eye bolts are provided at the top for flowmeters sized 8" (200mm) or    above, and further, a roll-prevention base is provided for flowmeters sized      ASME B 16.5 class 150 dimensions:  JIS B 2220 10K dimensions:  Meter size L1 L2 L3 No. of Weight  Meter size L1 (L2) L3 No. of Weight  (inch) (inch) (inch) (inch) bolts (lbs)  (mm) (mm) (mm) (mm) bolts (kg)  1/2 7.90 6.77  8.54  4 approx. 9.0  15 200 170 218 4 approx. 4.5  7.90  1 7.01
                    
                    ページ9に含まれる内容の要旨 
                    
                        TIC-LF654L                                                                                        Unit: inch (mm)  Weight: Approx. 7 lb (3.5 kg)  Note: Cable glands are not provided for LF622F cFMus approved type. Refer to the part Cable connection port at detector.  Note: 1 inch = 25.4 mm    Figure 7. Separate type converter LF622 and LF622F     External Connections  ● Combined type LF650/LF620 flowmeter and LF654/LF620F flowmeters      Instrument panel : Ordered separately Grounding with 100 
                    
                    ページ10に含まれる内容の要旨 
                    
                        TIC-LF654L    ● Separate type LF650/LF622 flowmeter and LF654/LF622F flowmeter  Instrument panel : Ordered separately  Grounding with 100  or less  2  wire 5.5mm or more  ground resistance  Power switch  Grounding with 100  or less  (External double-pole power switch)  ground resistance  Power supply  Thick walled steel conduit Current output (4  20mAdc)  Signal cable  or PROFIBUS  (2-wire shielded hard-rubber sheathed cable) Digital input  (20  30Vdc)  Excitation cable  or Modbus (3-wire shield
                    
                    ページ11に含まれる内容の要旨 
                    
                        TIC-LF654L    (6) The electromagnetic flowmeter is not equipped with   Wiring Precautions  terminating resistors. Use the terminating resistor  (1) Explosion proof type flowmeters are not provided  unit for PROFIBUS-PA or junction box, if  with cable glands.  necessary.  Refer to the part Cable connection port at detector  (7) Only one PROFIBUS-PA cable goes through a  and converter.  cable gland of the Electromagnetic Flowmeter.  (2)  Connect the grounding wire (IV wire 5.5mm² or  Please use t
                    
                    ページ12に含まれる内容の要旨 
                    
                        TIC-LF654L  Table 3. Flow Rate and Flow velocity   Ordering Information  1.  When ordering the LF654 series flowmeters, refer   (SI unit)  3 to Tables 4 and 5 (Type Specification Codes).   Unit: m /h   Flow rate An entry must be made for each of the columns in  Size 3.0 12   each of these tables.  0.1 m/s 0.3 m/s 1.0 m/s  (mm) m/s m/s  2.  Fluid characteristics:  15 0.06362 0.1908  0.6361  1.908 7.634 (1) Type of fluid to be measured and its characteristics  25 0.1767 0.5301  1.767  5.301 21.21
                    
                    ページ13に含まれる内容の要旨 
                    
                        TIC-LF654L  Table 4.  Specification Code (Flange type detector LF654 Series)  Model Specification Code  Detector category  Description  1 2 3 4 5 6 7 8 9 10 11 12 13 14 Gr.-A Gr.-B  L F 6 5         Flange type electromagnetic flowmeter detector     Usage                4 cFMus Class 1 Division 2 Hazardous Location √ √             Meter size        D       ½"(15mm) √        E       1"(25mm) √        S       1¼"(32mm) √        F       1½"(40mm) √        G       2"(50mm) √        W       2½"(65mm) 
                    
                    ページ14に含まれる内容の要旨 
                    
                        TIC-LF654L    Table 5.  Specification Code for converters  Model Specification Code  Contents LF620 LF622  1 2 3 4 5 6 7 8 9 1011 12 13 14  type type  L F 6 2          Electromagnetic flowmeter converter      0         Combined (Integral) type   –      2         Separate (Remote) type –                Purpose    F cFMus class I,  Division 2 approved                              Shape   A Standard type with case                 Converter mounting fitting    A None    –   C Panel, Acce