114, 115, 116, and 117 
Digital Multimeters 
Calibration Information 
Introduction 
XWWarning 
To avoid electric shock or injury, do not perform the performance tests or 
calibration adjustment procedures unless qualified to do so. 
The information provided in this document is for the use of qualified personnel 
only. 
The 114, 115, 116,  & 117 Calibration Information provides the information necessary to adjust and verify the 
performance of the Fluke Models 114, 115, 115C, 116, 116C, 117, 
                    
                    Streszczenie treści zawartej na stronie nr. 2 
                    
                        114, 115, 116, and 117  Calibration Information  Safety Information  "Warning" and "Caution" Statements    A “ XW Warning" identifies hazardous conditions and actions that could cause bodily harm or death.  A " W Caution" identifies conditions and actions that could damage the Meter, the equipment under test, or cause  permanent loss of data.  XWWarnings and Precautions   To avoid possible electric shock or personal injury, follow these guidelines:   • Use the Meter only as specified in this man
                    
                    Streszczenie treści zawartej na stronie nr. 3 
                    
                         Digital Multimeters   International Electrical Symbols   International Electrical Symbols  Table 1 lists the international symbols that appear in this document and on the Meter.  Table 1. Electrical Symbols  Symbol Description Symbol Description  B AC (Alternating Current) I Fuse  F DC (Direct Current) T Double Insulated  X Hazardous voltage W Important Information; Refer to manual  N Battery (Low battery when shown on the  J Earth ground  display.)  ~ Do not dispose of this product as unsorted
                    
                    Streszczenie treści zawartej na stronie nr. 4 
                    
                        114, 115, 116, and 117  Calibration Information  Accuracy Specifications  Accuracy  [1] Function Range Resolution Model  ± ([% of Reading] + [Counts])  DC millivolts 600.0 mV 0.1 mV 0.5 % + 2 114, 115, 116, 117 6.000 V 0.001 V  DC Volts  60.00 V 0.01 V 0.5 % + 2 114, 115, 116, 117 600.0 V 0.1 V   DC, 45 to 500 Hz 500 Hz to 1 kHz   [2] Auto-V LoZ True-rms 600.0 V 0.1 V 2.0 % + 3 4.0 % + 3 114, 116, 117   45 to 500 Hz 500 Hz to 1 kHz   [2] AC millivolts True-rms 600.0 mV 0.1 mV 1.0 % + 3 2.0 % + 3
                    
                    Streszczenie treści zawartej na stronie nr. 5 
                    
                         Digital Multimeters   Basic Maintenance   Accuracy Specifications (cont)  Accuracy  [1] Function Range Resolution Model  ± ([% of Reading] + [Counts])  99.99 Hz 0.01 Hz  999.9 Hz 0.1 Hz  [3] Hz (V or A input) 0.1 % + 2  115, 116, 117  9.999 kHz 0.001 kHz  50.00 kHz 0.01 kHz  Notes:  [1] Models listed in this column also refer to the “C” version of the model. For example, those rows containing model 115 are applicable to the 115C  as well.  [2] All ac ranges except Auto-V LoZ are specified from 
                    
                    Streszczenie treści zawartej na stronie nr. 6 
                    
                        114, 115, 116, and 117  Calibration Information  VoltAle rt 117 TRUE  RMS  MU LTIMETER OK  <.5 Lo / H HOLD MIN MAX RANGE OK Hz AUTO-V OFF LoZ V V mV Volt Alert A A Hz V A COM 10 A FUSED   erc010f.emf  Figure 1. Fuse Testing  Replacing the Battery and Fuse  XWWarning  To avoid shock, injury, or damage to the Meter:  • Remove test leads from the Meter before opening the case or battery door.  • Use ONLY a fuse with the amperage, interrupt voltage, and speed ratings  specified.  Battery Fuse   erc0
                    
                    Streszczenie treści zawartej na stronie nr. 7 
                    
                         Digital Multimeters   Performance Tests   2. Remove the Meter from its holster.  3. Remove two screws from the case bottom.  4. Separate the case bottom from the case top.  5. Remove the fuse from its holder and replace with an 11 A, 1000 V, FAST fuse having a minimum  interrupt rating of 17,000 A. Use only Fluke PN 803293.  6. To re-assemble the Meter, attach the case bottom to the case top and secure with the two screws. Insert  the Meter into its holster.  Cleaning the Meter  Wipe the case w
                    
                    Streszczenie treści zawartej na stronie nr. 8 
                    
                        114, 115, 116, and 117  Calibration Information  Table 2. Required Equipment (cont)  Recommended Equipment Measurement Function Accuracy  5500A Multi-product Calibrator (or Frequency (115, 116, and 117) 2 V, 50 kHz  equivalent) ±0.025 %  Fluke 80 AK  Temperature (116)   K-type Thermocouple Adapter Accessory  K-type Thermocouple, mini-plug on both Temperature (116)   ends  Double Banana plug VoltAlert (117)   Testing the Display  Push K and turn the rotary switch to the e position. Compare the di
                    
                    Streszczenie treści zawartej na stronie nr. 9 
                    
                         Digital Multimeters   Performance Tests   2. Warm up the calibrator as required by its specifications.  3. Allow the temperature of the UUT to stabilize at room temperature (23 °C ± 5 °C [73 °F  ± 9 °F] ).  4. Check the fuses and Battery, and replace them if necessary. Refer to “Testing the Fuses”, and  “Replacing the Battery and Fuse”.  To verify the accuracy of the DMM functions, do the following:  1. Connect the Calibrator to the VΩ and COM input terminals on the Meter.  2. Turn the rotary s
                    
                    Streszczenie treści zawartej na stronie nr. 10 
                    
                        114, 115, 116, and 117  Calibration Information  Table 3. DMM Performance Tests (cont)  Display Reading  Step Function Range Applied  [1] [1] [1] [1] 114 115 116 117  18 L  6.000V 0V -0.002 to 0.002  DC Volts  19 6.000V 5V 4.973 to 5.027  20 60.00V 50V 49.73 to 50.27  21 600.0V 600V 596.8 to 603.2  22 600.0V -600V -596.8 to -603.2  23 m 600.0 mV 6 mV, 45 Hz 5.6 to 6.4  AC Millivolts  24 600.0 mV 600mV, 1 kHz 587.7 to 612.3  25 •  600.0 mV 10 mV 9.7 to 10.3  DC Millivolts  26 600.0 mV 600mV 596.8
                    
                    Streszczenie treści zawartej na stronie nr. 11 
                    
                         Digital Multimeters   Performance Tests   Table 3. DMM Performance Tests (cont)  Display Reading  Step Function Range Applied  [1] [1] [1] [1] 114 115 116 117  41 VoltAlert Hi  Refer to  steps 1 – 5  N/A N/A N/A in the  procedure  below  42 VoltAlert Lo  Refer to  steps 6 – 9  N/A N/A N/A in the  procedure  below  [1] If using a Fluke 9100 calibrator, the Calibrator Frequency mode must be used to obtain accurate frequency.  [2] To keep from tripping the calibrator to standby, ramp up the voltag
                    
                    Streszczenie treści zawartej na stronie nr. 12 
                    
                        VoltAlert TRUE  RMS  MULTIMETER 117 TRIG STBY 7 8 4 5 . 2 +/- 0 . SCOPE OUT AUX A SENSE AUX V NORMAL V RTD 114, 115, 116, and 117  Calibration Information  Insert Double Banana Center Meter top on Hi terminal. Note that LED is RED   erc013f.emf  Figure 4. VoltAlert Testing  8. Hold the Meter so that the top is vertically and horizontal centered to the banana plug’s Hi terminal.   Verify that the Meter’s beeper is sounding continuously and the red LED at the top of the display lights.  9. Return 
                    
                    Streszczenie treści zawartej na stronie nr. 13 
                    
                         Digital Multimeters   Calibration Adjustment   2. Press g once to see the calibration counter.   3. Press g again to start the password entry. The Meter displays “ ?>>>”  4. The Meter buttons indicated below represent the numbers 1 through 5 when entering or changing the  password:  K = 1 M = 2 q = 3 g = 4 Q = 5    5. Press 4 buttons to enter the current password. If changing the password for the first time, enter  K (1), M (2), q (3), and g (4).  6. Press qto change the password. The Meter dis
                    
                    Streszczenie treści zawartej na stronie nr. 14 
                    
                        114, 115, 116, and 117  Calibration Information  Meter Buttons Used in the Calibration Steps  When performing the Calibration Adjustment Procedure, the Meter buttons behave as follows. This may be of help  determining why a calibration step is not accepted and for determining the input value without referring to Table 4.  Press and hold K to show the measured value. The measured value is not calibrated so it may not match the  input value. This is normal.  Press and hold M to display the require
                    
                    Streszczenie treści zawartej na stronie nr. 15 
                    
                         Digital Multimeters   Calibration Adjustment   Notes  Set the calibrator to Standby prior to changing the function switch position and after  completing adjustment of each function.   If the calibration adjustment procedure is not properly completed, the Meter will not operate  correctly.  Table 4. Calibration Adjustment Steps  Rotary Calibration Steps  Input Calibrator  Switch  [1] [1] [1,2] Terminals Source Value  114 115 116 117  Position  [2] e N/A C/01 C/01 C/01 No leads No leads  Ohms  m 
                    
                    Streszczenie treści zawartej na stronie nr. 16 
                    
                        114, 115, 116, and 117  Calibration Information  Table 4. Calibration Adjustment Steps (cont)  Rotary Calibration Steps  Input Calibrator  Switch  [1] [1] [1,2] Terminals Source Value  114 115 116 117  Position  $ N/A N/A C/23 N/A + and COM 600 μA, 0 Hz  DC μamps  [1] Models listed in this column also refer to the “C” version of the model. For example, model 115 steps are valid for the 115C.  [2] Do not calibrate the 117 or 117C with a line-frequency power source nearby (e.g. fluorescent light, 
                    
                    Streszczenie treści zawartej na stronie nr. 17 
                    
                         Digital Multimeters   Replacement Parts   Replacement Parts  Table 5 lists the Meter’s replacable parts identified in Figure 6.  6 5 8 12 3 9 1 10 4 2 7 11 15 4 PL 23 13 16 17 2 PL 19 20 14 18 21 22   erc14.emf  Figure 6. Exploded View of Meter    17                                                                                                                                                                                                                                                         
                    
                    Streszczenie treści zawartej na stronie nr. 18 
                    
                        114, 115, 116, and 117  Calibration Information  Table 5. Replacable Parts List  Part  Item Description Qty.  Number  1 LCD,FLUKE-11X,3.2V,TN,4-DIGIT,1/4-DUTY,1/3-BIAS,LEPTON 2509955 1  2 CONNECTOR,ELASTOMERIC,.010 IN CTR,.218 IN HIGH,.090 IN 1  THK,2.284 IN LONG,BULK 2534229  3 FLUKE-117-2006,BRACKET MASK, 117 2525608 1  FLUKE-117-2006-06,BRACKET MASK, 117 China 2631059 1  FLUKE-117-2006-01,BRACKET MASK, 114 2527431 1  FLUKE-117-2006-02,BRACKET MASK, 115 2527446 1  FLUKE-117-2006-04,BRACKET MAS
                    
                    Streszczenie treści zawartej na stronie nr. 19 
                    
                         Digital Multimeters   Warranty   Table 5. Replaceable Parts List  Part  Item Description Qty.  Number  19 BATTERY,PRIMARY,MNO2- 1  ZN,9V,505MAH,6LR61,ALKALINE, 17X26X48MM,BULK 614487  20 FLUKE 12-8004,SHOCK ABSORBER 878983 1  21 FLUKE-117-2005,TILT STAND 2525594 1  FLUKE-117-2005,TILT STAND, 11X China 2631071 1  22 FLUKE-117-2010,HOLSTER 2525649 1  FLUKE-117-2010,HOLSTER, 11X China 2631080 1  23 FUSE,11A,1000V,FAST.406INX1.5IN,BULK 803293 1 (115, 117)  Not shown MANUAL,114/115/117 USERS MANUAL,
                    
                    Streszczenie treści zawartej na stronie nr. 20 
                    
                        114, 115, 116, and 117  Calibration Information    20