                        Owner’s Operator and Maintenance Manual
MA50 Series
MA50 Alternating Pressure 
Overlay System
MA51 Alternating Pressure 
Mattress System
MA55 Alternating Pressure  
Low Air Loss Mattress System
DEALER: This manual MUST be given to 
the user of the product.
USER: BEFORE using this product, read this 
manual and save for future reference.
For more information regarding 
Invacare products,  parts, and services, 
please visit www.invacare.com                                                          
                    
                    Краткое содержание страницы № 2 
                    
                        SPECIAL NOTES  WARNING DO NOT use this product or any available optional equipment without first  completely reading and understanding these instructions and any additional  instructional material such as owner’s manuals, service manuals or instruction  sheets supplied with this product or optional equipment. If you are unable to  understand the warnings, cautions or instructions, contact a healthcare  professional, dealer or technical personnel before attempting to use this equipment  - otherw
                    
                    Краткое содержание страницы № 3 
                    
                        TABLE OF CONTENTS TABLE OF CONTENTS SPECIAL NOTES ................................................................................2 REGISTER YOUR PRODUCT ............................................................... 4 TYPICAL PRODUCT PARAMETERS .................................................... 5 Electrical Parameters .................................................................................................................................5 Performance Parameters ......................
                    
                    Краткое содержание страницы № 4 
                    
                        TABLE OF CONTENTS TABLE OF CONTENTS SECTION 3—MAINTENANCE AND TROUBLESHOOTING ................... 21 Cleaning the System.................................................................................................................................21 Storing the System...............................................................................................................................22 Troubleshooting.....................................................................................
                    
                    Краткое содержание страницы № 5 
                    
                        TYPICAL PRODUCT PARAMETERS TYPICAL PRODUCT PARAMETERS Electrical Parameters MA50/MA51/MA55 INPUT VOLTAGE AC: 90 V INPUT FREQUENCY: 60 Hz CURRENT: 1 A MAXIMUM POWER CONSUMPTION: 5 ± 3 W CIRCUIT PROTECTION: Dual fused, 250 V, 1 A fast blow fuses MODE OF OPERATION: Continuous Performance Parameters MA50/MA51/MA55 WEIGHT CAPACITY STANDARD MATTRESS: 350 lbs BARIATRIC MATTRESS: 1000 lbs PRESSURE ZONE: 2 MAXIMUM FLOW: 8 ± 4 LPM MAXIMUM FLOW PRESSURE: 35 ± 5 mmHg MAXIMUM FLOW TIMER: 45 minutes SUPPORT S
                    
                    Краткое содержание страницы № 6 
                    
                        TYPICAL PRODUCT PARAMETERS Environmental Parameters MA50/MA51/MA55 OPERATING CONDITIONS AMBIENT TEMPERATURE: 50° - 104° F RELATIVE HUMIDITY: 30% - 75% Non-Condensing ATMOSPHERIC PRESSURE: 70 - 106 kPa STORAGE AND SHIPPING CONDITIONS AMBIENT TEMPERATURE: -40° - 158° F RELATIVE HUMIDITY: 10% - 100%  Non-Condensing ATMOSPHERIC PRESSURE: 50 - 106 kPa MA50 Series 6 Part No 1148136                                                                                                                          
                    
                    Краткое содержание страницы № 7 
                    
                        LABEL LOCATION LABEL LOCATION DANGER-EXPLOSION HAZARD: DO NOT use in the presence of  flammable anesthetics. CAUTION: Equipment should be connected to a properly grounded  receptacle (3-prong). Risk of Electrical shock. DO NOT remove  back. Disconnect air hose before administering CPR. NOTE: This label is on the back of  the unit and also contains the  serial number.  WARNING Patient entrapment with bed side rails may cause injury or death. Mattress  MUST fit bed frame and side rails snugly to 
                    
                    Краткое содержание страницы № 8 
                    
                        SECTION 1—GENERAL GUIDELINES SECTION 1—GENERAL GUIDELINES  WARNING SECTION 1 - GENERAL GUIDELINES contains important information for the safe  operation and use of this product. Contraindications ALWAYS consult the patient’s physician before using the MA50, MA51 and MA55 systems. Installation The MA50, MA51 and MA55 systems MUST be installed on medical bed frames with side  rails. The side rails MUST be in the raised position whenever a patient is on the bed. Controls on the footboard may be ob
                    
                    Краткое содержание страницы № 9 
                    
                        SECTION 1—GENERAL GUIDELINES Fire Hazard  DANGER Smoking DO NOT SMOKE while using this device. This system uses room air for circulation  through the mattress. A cigarette can burn a hole in the bed surface and cause  damage to the mattress. Also, patient clothing, bed sheets, etc. may be combustible  and cause a fire. Failure to observe this warning can result in severe fire, property  damage and cause physical injury or death. Smoking by visitors in the room will contaminate the system. There
                    
                    Краткое содержание страницы № 10 
                    
                        SECTION 2—OPERATION SECTION 2—OPERATION CAUTION The control unit and mattress on the MA50 series are designed to be used as a  system. DO NOT replace mattresses or control units with other models or other  brands. Otherwise, damage to the system may occur. Contact your supplier to get  the correct replacement if needed. Installing the Mattress Overlay System (MA50) NOTE: For this procedure, refer to FIGURE 2.1. NOTE: The overlay system comes with the following straps - head end elastic strap, fo
                    
                    Краткое содержание страницы № 11 
                    
                        SECTION 2—OPERATION Mattress Replacement System (MA51 and MA55) CAUTION DO NOT strap the mattress to the bed frame at the head and foot ends. Secure  ALL mattress straps. Secure the straps to the bed deck at the head and foot ends  and to the frame at the center of the bed. Otherwise damage to the mattress will  occur when the head and foot ends are raised. Head End Bed Deck Center of Bed Frame Foot End Bed Deck DO NOT Strap Here NOTE: For this procedure, refer to FIGURE 2.2. NOTE: The powered m
                    
                    Краткое содержание страницы № 12 
                    
                        SECTION 2—OPERATION Installing the Side Rails  WARNING Patient entrapment with bed side rails may cause injury or death. Mattress MUST  fit bed frame and side rails snugly to prevent patient entrapment. Follow the  manufacturer’s instructions. Monitor patient frequently. Read and understand the  Owner’s/Operator’s Manual prior to using this equipment. Invacare product  manuals are available at www.invacare.com or your dealer. NOTE: Refer to the instructions provided with the side rails for the 
                    
                    Краткое содержание страницы № 13 
                    
                        SECTION 2—OPERATION Control Unit Connectors Tab Hose  Connector Hose FIGURE 2.4   Connecting the Hose Connecting the Power Cord  WARNING DO NOT alter plug to fit a non-conforming outlet. Instead, have an electrician  install a properly grounded 3-prong outlet. Failure to use the correct plug and outlet  can result in a potential safety hazard. CAUTION Ensure that the power cord of the control unit is not pinched, or has any objects  placed on it, and also ensure it is not located where it can b
                    
                    Краткое содержание страницы № 14 
                    
                        SECTION 2—OPERATION Using the Front Panel NOTE: For this procedure, refer to FIGURE 2.5. DETAIL “A” - MA50 AND MA51 FRONT PANEL Power Fail LED Low Pressure  LED Max Inflate  Fowler  Button Button Soft Button Firm Button Alarm Silence  Static/Alternating  Power  LCD  Lock Button Button Pressure Button Button Display DETAIL “B” - MA55 FRONT PANEL Power Fail LED Low Pressure  LED Mode Button  Fowler  (Max Inflate/Low  Button Air Loss) Soft Button Firm Button Alarm Silence  LCD  Static/Alternating  
                    
                    Краткое содержание страницы № 15 
                    
                        SECTION 2—OPERATION Static/Alternating Pressure Button 1. Press the Alternating Pressure button ( ) to select between Static Mode and one  of four Alternating Pressure Times. NOTE: The alternating pressure times indicate the frequency of deflation or inflation of half of the  air cushions (even or odd numbered). For example, 10 minutes is selected. In this example, the even  numbered air cushions in the mattress will deflate, while the odd numbered cushions remain at  constant pressure. When 10 
                    
                    Краткое содержание страницы № 16 
                    
                        SECTION 2—OPERATION Firm/Soft Buttons 1. Select comfort pressure settings by pressing Firm ( ) or Soft ( ) button. • Soft Button - Pressing this button reduces the pressure setting in the mattress. • Firm Button - Pressing this button increase the pressure setting in the mattress. NOTE: The patient comfort pressure ranges from Soft (level 0 = 6±4 mmHg) to Firm (level 9 =  32±6 mmHg). The Comfort Control LED displays the patient comfort pressure levels from 0 to 9  and provides a guide to the car
                    
                    Краткое содержание страницы № 17 
                    
                        SECTION 2—OPERATION Mode Button (Max Inflate/Low Air Loss) NOTE: The Mode button is on the MA55 system only. 1. Press the Mode button ( ) to select the Max Inflate mode or the Low Air Loss  mode. • Max Inflate Mode - In this mode, the mattress inflates rapidly to maximum  firmness (pressurized to 35 ± 6 mmHg). A series of beeps sound every three  minutes as a reminder that the Max Inflate mode is active. After 45 minutes, the  Max Inflate mode deactivates and the control unit defaults to the pre
                    
                    Краткое содержание страницы № 18 
                    
                        SECTION 2—OPERATION Low Pressure LED In the event that the mattress hose disconnects, an alarm sounds and Low Pressure LED  flashes AMBER. Once the low pressure problem is fixed, the control unit resumes  operation in the previously set mode. Powering Up the System NOTE: For this procedure, refer to FIGURE 2.5 on page 14. 1. Turn on the power to the system by pressing the Power button on the control unit. NOTE: Once the button is released, a GREEN LED illuminates when the unit is on. Placing the
                    
                    Краткое содержание страницы № 19 
                    
                        SECTION 2—OPERATION 9. Verify that the patient has not bottomed out by performing the following steps: A. Ensure that the patient is lying flat on his/her back in the middle of the mattress. B. Place four fingers between the air cushions directly underneath the sacral region  of the patient’s body. C. Ensure that there is 3 to 4-finger width clearance between the bottom of the patient  and the bed frame. D. Adjust the comfort setting, if needed. E. Wait five minutes for the mattress pressure to 
                    
                    Краткое содержание страницы № 20 
                    
                        SECTION 2—OPERATION Transferring Patient From/To a Wheelchair  WARNING ALWAYS engage the wheel locks of the bed and the wheel locks of the wheelchair  before transferring the patient between the bed and the wheelchair. 1. Engage the wheel locks of the bed. Refer to the owner’s manual provided with the bed. 2. Engage the wheel locks of the wheelchair, if applicable. Refer to the owner’s manual  provided with the wheelchair. 3. Press the Max Inflate or Mode button to achieve maximum mattress pres