Fluke 9140 user manual

User manual for the device Fluke 9140

Device: Fluke 9140
Category: Home Safety Product
Manufacturer: Fluke
Size: 0.53 MB
Added : 1/23/2014
Number of pages: 52
Print the manual

Download

How to use this site?

Our goal is to provide you with a quick access to the content of the user manual for Fluke 9140. Using the online preview, you can quickly view the contents and go to the page where you will find the solution to your problem with Fluke 9140.

For your convenience

If looking through the Fluke 9140 user manual directly on this website is not convenient for you, there are two possible solutions:

  • Full Screen Viewing - to easily view the user manual (without downloading it to your computer), you can use full-screen viewing mode. To start viewing the user manual Fluke 9140 on full screen, use the button Fullscreen.
  • Downloading to your computer - You can also download the user manual Fluke 9140 to your computer and keep it in your files. However, if you do not want to take up too much of your disk space, you can always download it in the future from ManualsBase.
Fluke 9140 User manual - Online PDF
Advertisement
« Page 1 of 52 »
Advertisement
Print version

Many people prefer to read the documents not on the screen, but in the printed version. The option to print the manual has also been provided, and you can use it by clicking the link above - Print the manual. You do not have to print the entire manual Fluke 9140 but the selected pages only. paper.

Summaries

Below you will find previews of the content of the user manuals presented on the following pages to Fluke 9140. If you want to quickly view the content of pages found on the following pages of the manual, you can use them.

Abstracts of contents
Summary of the content on the page No. 1

Hart Scientific
9140
Dry-wellCalibrator
User’sGuide
Rev. 5B1702

Summary of the content on the page No. 2

Fluke Corporation, Hart Scientific Division 799 E. Utah Valley Drive • American Fork, UT 84003-9775 � USA Phone: +1.801.763.1600 � Telefax: +1.801.763.1010 E-mail: support@hartscientific.com www.hartscientific.com Subject to change without notice. � Copyright © 2005 � Printed in USA Rev. 5B1702

Summary of the content on the page No. 3

Table of Contents 1 Before You Start . . . . . . . . . . . . . . . . . . . . . . . . . . 1 1.1 Symbols Used . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 1.2 Safety Information . . . . . . . . . . . . . . . . . . . . . . . . . . 2 1.2.1 WARNINGS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 1.2.2 CAUTIONS . . . . . . . . . . . . . . .

Summary of the content on the page No. 4

7.2 Temperature Set-point . . . . . . . . . . . . . . . . . . . . . . . . 21 7.2.1 Programmable Set-points . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21 7.2.2 Set-point Value . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23 7.2.3 Temperature Scale Units . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23 7.3 Scan . . . . . . . . . . . .

Summary of the content on the page No. 5

10 Calibration Procedure . . . . . . . . . . . . . . . . . . . . . . 43 10.1 Calibration Points . . . . . . . . . . . . . . . . . . . . . . . . . . 43 10.2 Calibration Procedure . . . . . . . . . . . . . . . . . . . . . . . . 43 10.2.1 Compute DELTA: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43 10.2.2 Compute R0 & ALPHA: . . . . . . . . . . . . . . . . . . .

Summary of the content on the page No. 6

Figures Figure 1 9140 Back Panel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13 Figure 2 9140 Front Panel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14 Figure 3 Inserts available for the 9140 block assembly . . . . . . . . . . . . . . 15 Figure 4 Controller Operation Flowchart . . . . . . . . . . . . . . . . . . . . . 22 Figure 5 Well temperature fluctuation at various proportion

Summary of the content on the page No. 7

Tables Table 1 International Electrical Symbols . . . . . . . . . . . . . . . . . . . . . 1 Table 2 9140 controller communications commands. . . . . . . . . . . . . . . 36 Table 3 9140 controller communications commands continued . . . . . . . . . 37 v

Summary of the content on the page No. 8

1 Before You Start Symbols Used 1 Before You Start 1.1 Symbols Used Table 1 lists the symbols used on the instrument or in this manual and the meaning of each symbol. Table 1 International Electrical Symbols Symbol Description AC (Alternating Current) AC-DC Battery Complies with European Union directives DC Double Insulated Electric Shock Fuse PE Ground Hot Surface (Burn Hazard) Read the User’s Manual (Important Information) Off On 1

Summary of the content on the page No. 9

9140 Dry-well Calibrator User’s Guide Symbol Description Canadian Standards Association OVERVOLTAGE (Installation) CATEGORY II, Pollution Degree 2 per IEC1010-1 re- fers to the level of Impulse Withstand Voltage protection provided. Equipment of OVERVOLTAGE CATEGORY II is energy-consuming equipment to be supplied from the fixed installation. Examples include household, office, and laboratory appliances. C-TIC Australian EMC The European Waste Electrical and Electronic Equipment (WEEE) Directive

Summary of the content on the page No. 10

1 Before You Start Safety Information ing power such as storage in a low humidity temperature chamber operat- ing at 50°C for 4 hours or more. • DO NOT use this instrument for any application other than calibration work. The instrument was designed for temperature calibration. Any other use of the instrument may cause unknown hazards to the user. • Completely unattended operation is not recommended. • Overhead clearance is required. DO NOT place the instrument under a cabinet or other structure.

Summary of the content on the page No. 11

9140 Dry-well Calibrator User’s Guide ELECTRICAL SHOCK • These guidelines must be followed to ensure that the safety mechanisms in this instrument will operate properly. This instrument must be plugged into a 115 VAC, 60Hz (230 VAC, 50Hz optional), AC only electric outlet. The power cord of the instrument is equipped with a three-pronged grounding plug for your protection against electrical shock hazards. It must be plugged directly into a properly grounded three-prong receptacle. The receptacle

Summary of the content on the page No. 12

1 Before You Start Authorized Service Centers • The instrument and any thermometer probes used with it are sensitive in- struments that can be easily damaged. Always handle these devices with care. DO NOT allow them to be dropped, struck, stressed, or overheated. • The Factory Reset Sequence (see Section 12.1, Troubleshooting) should be performed only by authorized personnel if no other action is successful in correcting a malfunction. You must have a copy of the most recent Re- port of Calibrat

Summary of the content on the page No. 13

9140 Dry-well Calibrator User’s Guide Science Park Eindhoven 5108 5692 EC Son NETHERLANDS Phone: +31-402-675300 Telefax: +31-402-675321 E-mail: ServiceDesk@fluke.nl Fluke Int'l Corporation Service Center - Instrimpex Room 2301 Sciteck Tower 22 Jianguomenwai Dajie Chao Yang District Beijing 100004, PRC CHINA Phone: +86-10-6-512-3436 Telefax: +86-10-6-512-3437 E-mail: xingye.han@fluke.com.cn Fluke South East Asia Pte Ltd. Fluke ASEAN Regional Office Service Center 60 Alexandra Terrace #03-16 The C

Summary of the content on the page No. 14

2 Introduction 2 Introduction The Hart Scientific 9140 Mid-Range Field Calibrator may be used as a portable instrument or bench top temperature calibrator for calibrating thermocouple and RTD temperature probes. The 9140 is small enough to use in the field, and accurate enough to use in the lab. The instrument features: • Rapid heating and cooling • Interchangeable multiple hole aluminum probe sleeves • Convenient hand strap • RS-232 interface capability Built in programmable features include: •

Summary of the content on the page No. 15

3 Specifications and Environmental Conditions Specifications 3 Specifications and Environmental Conditions 3.1 Specifications The following table lists the specifications for this instrument. Accuracy speci- fications are applicable for a one-year calibration interval. In line with normal prudent metrology practices, Hart Scientific recommends a short-cycle interval of six months for new units during the first year. Power 115 VAC (±10%), 4.4 A or 230 VAC (±10%), 2.2 A, switchable 500 W Ambient T

Summary of the content on the page No. 16

9140 Dry-well Calibrator User’s Guide operated in an excessively dusty or dirty environment. Maintenance and clean- ing recommendations can be found in the Maintenance Section of this manual. The instrument operates safely under the following conditions: • temperature range: 5–50°C (41–122°F) • ambient relative humidity: 15–50% • pressure: 75kPa–106kPa • mains voltage within ±10% of nominal • vibrations in the calibration environment should be minimized • altitude does not effect the performance

Summary of the content on the page No. 17

4 Quick Start Unpacking 4 Quick Start 4.1 Unpacking Unpack the dry-well carefully and inspect it for any damage that may have oc- curred during shipment. If there is shipping damage, notify the carrier immediately. Verify that the following components are present: • 9140 Dry-well • Insert • Insert Removal Tool • Power Cord • Serial Cable • User's Guide • 9930 Software Package 4.2 Set-Up Place the calibrator on a flat surface with at least 6 inches of free space around the instrument. Plug the po

Summary of the content on the page No. 18

9140 Dry-well Calibrator User’s Guide dry-well will turn on and begin to heat to the previously programmed tempera- ture set-point. The front panel LED display will indicate the actual dry-well temperature. 4.4 Setting the Temperature Section 7.2 explains in detail how to set the temperature set-point on the cali- brator using the front panel keys. The procedure is summarized here. (1) Press “SET” twice to access the set-point value. (2) Press “UP” or “DOWN” to change the set-point value. (3) Pr

Summary of the content on the page No. 19

5 Parts and Controls Rear Panel 5 Parts and Controls The user should become familiar with the dry-well calibrator and its parts: 5.1 Rear Panel Figure 1 on page 13. Power Cord - At the rear of the calibrator is the removable power cord inlet that plugs into an IEC grounded socket. Power Switch - The power switch is located on the power entry module (PEM). The PEM also houses the fuses and the dual voltage selector. The PEM and Heater Voltage Switch (see below) allow the unit to be field switchab

Summary of the content on the page No. 20

9140 Dry-well Calibrator User’s Guide around the two corners of the calibrator are provided for airflow. The area around the calibrator must be kept clear to allow adequate ventilation. The air- flow is directed upward and as a result, can be extremely hot. 5.2 Front Panel Figure 2 on page 14. Controller Display - The digital display is an important part of the temperature controller because it not only displays set and actual temperatures but also vari- ous calibrator functions, settings, and c


Alternative user manuals
# User manual Category Download
1 Fluke Ti20 User manual Home Safety Product 78
2 Sony HVL-F5DF User manual Home Safety Product 2
3 Sony HVL-HFL1 User manual Home Safety Product 6
4 Sony HVL-FDH2 User manual Home Safety Product 3
5 ACDelco LG08 User manual Home Safety Product 3
6 Adaptec 2000 Series Drive-Thru Timer System 2+2/386 User manual Home Safety Product 1
7 American Dynamics 1461 User manual Home Safety Product 1
8 AMX Video Sync Sensor VSS2 User manual Home Safety Product 0
9 AOSafety Beyond The Standard Eyewear Lens User manual Home Safety Product 1
10 AOSafety Eye Gear Lexa User manual Home Safety Product 0
11 AOSafety Eye Gear Nassau Vibrance User manual Home Safety Product 1
12 AOSafety GoggleGear Dust User manual Home Safety Product 1
13 AOSafety Nassau Rave ANSI Z87.1 User manual Home Safety Product 13
14 AOSafety Eyeware Zora 32150 User manual Home Safety Product 1
15 AOSafety Polarized Eyewear GlareX User manual Home Safety Product 1